Mallinckrodt Manufacturing LLC FDA Inspection Record

FEI 3004531588 · Inspections, Form 483s, warning letters and import refusals

Mallinckrodt Manufacturing LLC is an FDA-registered facility (FEI 3004531588) located in MADISON, WI, US. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Mar 13, 2015: No Action Indicated (NAI)
Inspections on record
3 (2 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
5, latest Aug 8, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 13, 2015No Action Indicated (NAI)No citations listed
Oct 21, 2011Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Servicing - Lack of or inadequate procedures (21 CFR 820.200(a))
Report of risk to health (21 CFR 806.10(a)(1))
Records of communications (21 CFR 806.20(b)(5))
Aug 17, 2010Voluntary Action Indicated (VAI)Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Lack of or inadequate procedures (21 CFR 820.100(a))
Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))
Documentation (21 CFR 820.100(b))
and 1 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Aug 8, 2023Class IZ-2450-2023Mallinckrodt One-Way Valve, 22F x 22M, a) Part #91346 (1 piece) and b) Part #50412 (pack…
The One-Way Valve, 22F x 22M may stick and prevent or reduce the flow of ventilated air or oxygen.
Nov 21, 2016Class IIZ-1076-2017Mallinckrodt Pharmaceuticals INOmax DSIR Plus delivery system. Model 10007. Delivers…
Mallinckrodt discovered that some devices may have been manufactured with an out-of-tolerance manometer which can result in in an inaccurate…
Jan 14, 2015Class IIZ-1223-2015Ikaria, INOmax DSIR (Delivery System), Model 10007. Nitric oxide delivery system for use…
Potential delivery failure alarm condition. INOmax DSIR with software version 2.0.4 and a certain variant of the monitor display may…
Oct 17, 2014Class IIZ-0575-2015INOMAX DSIR Nitric Oxide delivery system, Model 10007, software version 3.1.0., delivers…
An issue has been identified in the INOmax DSIR system that could result in monitored Nitric Oxide (NO) concentration reporting lower than…
Mar 27, 2014Class IIZ-1528-2014INOmax DSIR, model 10007, Nitric Oxide Delivery System, with Software version 3.0.0. The…
An anomaly impacts the monitoring system of INOmax DSIR Plus (software 3.0.0). When the integrated pneumatic backup switch is activated…

FDA record score

9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 27, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-2450-2023, this FEI as the recalling firmAug 8, 202315 x 0.69
Class II recall Z-1076-2017, this FEI as the recalling firmNov 21, 201610 years or older, so it adds no points0
Class II recall Z-1223-2015, this FEI as the recalling firmJan 14, 201510 years or older, so it adds no points0
Class II recall Z-0575-2015, this FEI as the recalling firmOct 17, 201410 years or older, so it adds no points0
Class II recall Z-1528-2014, this FEI as the recalling firmMar 27, 201410 years or older, so it adds no points0
Inspection classified VAIOct 21, 201110 years or older, so it adds no points0
Inspection classified VAIAug 17, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 27, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 27, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 27, 2026Newest record Nov 19, 2026Current
Warning letter textNov 27, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 27, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Mallinckrodt Manufacturing LLC as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Mallinckrodt Manufacturing LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Mallinckrodt Manufacturing LLC → Check this supplier

Facility details

Company
Mallinckrodt Manufacturing LLC
Location
MADISON, WI (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.