FENWAL, INC. FDA Inspection Record

FEI 3004548776 · Inspections, Form 483s, warning letters and import refusals

FENWAL, INC. is an FDA-registered facility (FEI 3004548776) located in LAKE ZURICH, IL, US. RegDataLab tracks 2 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Aug 7, 2017: No Action Indicated (NAI)
Inspections on record
2 (1 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
1 listed by FDA for this FEI, latest Oct 14, 2016
Recalls as the recalling firm
13, latest May 28, 2025

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Aug 7, 2017No Action Indicated (NAI)No citations listed
Mar 20, 2009Voluntary Action Indicated (VAI)Investigation of device failures (21 CFR 820.198(c))
Changes to correct/prevent quality problems (21 CFR 820.100(a)(5))
Identification of actions needed (21 CFR 820.100(a)(3))

Recalls

Showing the 10 most recent of 13 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 28, 2025Class IIB-0394-2025ACD (Anticoagulant Citrate Dextrose Solution) SINGLE BPU PL-146 W/IDT, Product Code…
In this situation, Fenwal International (a Fresenius Kabi Company) is initiating a voluntary recall for one lot of Fenwal Blood-Pack Units…
May 27, 2025Class IIB-0379-2025CPD/ADSOL TRIPLE BIOFLEX SAMPLE DIVERSION PL146 PL2209, Product Code 4R1787, NDC number… and 1 more product
In this situation, Fenwal International (a Fresenius Kabi Company) is initiating a voluntary recall for one lot of Fenwal Blood-Pack Unit…
Feb 15, 2024Class IIB-0212-2024Fenwal Blood-Pack Units with Integral Donor Tube Using ACD or CPD and Fenwal HighFlo… and 8 more products
Fenwal International, Inc. (a Fresenius Kabi Company) is initiating a voluntary recall for certain lots of Fenwal Blood-Pack Unit product…
Feb 14, 2023Class IIZ-1254-2023Amicus MNC Apheresis Kit Double Needle. Component of Amicus Separator multiprocedural… and 1 more product
Potential for centrifuge packs to develop a stress leak for certain lots of Amicus MNC Apheresis kit and Amicus Exchange kit (Therapeutic…
Feb 25, 2021Class IIZ-1343-2021Amicus MNC Apheresis Kit - Double Needle - Product Usage: intended for use in the… and 2 more products
Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using…
Jun 24, 2019Class IZ-2064-2019VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and… and 1 more product
The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of…
Mar 5, 2019Class IIZ-1209-2019Fresenius Kabi CATSmart device (Continuous autotransfusion System). Labeled as the…
Based on internal investigations, Fresenius Kabi has identified a multifaceted set of events involving the CATSmart Systems, including the…
Feb 11, 2019Class IIZ-1059-2019Vacuum Unit for CATSmart Vacuum Pump. Fresenius CATSmart (Continuous Autotransfusion…
The optional Vacuum Unit may stop working and display the failure message "Failure vacuum unit" during use.
Apr 9, 2018Class IIIB-0602-2018DXT Relay 2.X Software
Fresenius Kabi's DXT Relay 2.X Software, with a defect or glitch, was distributed.
Sep 3, 2013Class IIIB-0076-14Blood Pack Unit with an Integrally Attached Container of ADSOL Red Cell Preservation…
Mislabeled Blood-Pack Units were distributed.

FDA record score

20 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

11 records behind this number
RecordDatePoints x age weightAdds
Class II recall B-0394-2025, this FEI as the recalling firmMay 28, 20254 x 14
Class II recall B-0379-2025, this FEI as the recalling firmMay 27, 20254 x 14
Class I recall Z-2064-2019, this FEI as the recalling firmJun 24, 201915 x 0.253.8
Class II recall B-0212-2024, this FEI as the recalling firmFeb 15, 20244 x 0.62.4
Class II recall Z-1254-2023, this FEI as the recalling firmFeb 14, 20234 x 0.62.4
Class II recall Z-1343-2021, this FEI as the recalling firmFeb 25, 20214 x 0.251
Class II recall Z-1209-2019, this FEI as the recalling firmMar 5, 20194 x 0.251
Class II recall Z-1059-2019, this FEI as the recalling firmFeb 11, 20194 x 0.251
Warning letter on FDA's dashboard, which does not show close-outsOct 14, 201610 years or older, so it adds no points0
Class II recall Z-1523-2013, this FEI as the recalling firmMay 24, 201310 years or older, so it adds no points0
Inspection classified VAIMar 20, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing FENWAL, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check FENWAL, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Fenwal Inc
Location
LAKE ZURICH, IL (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.