Janssen Diagnostics, LLC FDA Inspection Record

FEI 3004619490 · Inspections, Form 483s, warning letters and import refusals

Janssen Diagnostics, LLC is an FDA-inspected facility (FEI 3004619490) located in Raritan, NJ, US. RegDataLab tracks 1 FDA inspection at this facility.

What FDA found at this site

Latest inspection
Dec 17, 2008: Voluntary Action Indicated (VAI)
Inspections on record
1 (1 VAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
3, latest May 15, 2015

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Dec 17, 2008Voluntary Action Indicated (VAI)Verification or validation of changes (21 CFR 820.70(b))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 15, 2015Class IIZ-2406-2015Cell Search Circulating Tumor Kit. Intended for the enumeration of circulating tumor cells…
Complaints of the presence of artifacts that appears as small bead like structures in the image gallery when processing both control cells…
Jan 8, 2014Class IIZ-1195-2014CELLTRACKS ANALYZER II¿ (Product Code 9555) Product Usage: The CELLTRACKS ANALYZER…
Janssen Diagnostics, LLC received complaints of camera failures on the CELLTRACKS ANALYZER II¿.
Dec 28, 2012Class IIZ-1949-2013CELLTRACKS AUTOPREP System (Software Version 2.4.1 & Below. Product Usage: The… and 1 more product
Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID)…

FDA record score

0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 16, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

4 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2406-2015, this FEI as the recalling firmMay 15, 201510 years or older, so it adds no points0
Class II recall Z-1195-2014, this FEI as the recalling firmJan 8, 201410 years or older, so it adds no points0
Class II recall Z-1949-2013, this FEI as the recalling firmDec 28, 201210 years or older, so it adds no points0
Inspection classified VAIDec 17, 200810 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 16, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 16, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 16, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 16, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing Janssen Diagnostics, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Janssen Diagnostics, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Janssen Diagnostics, LLC
Location
Raritan, NJ (US)
Source
inspections