MAST BIOSURGERY USA INC. FDA Inspection Record
MAST BIOSURGERY USA INC. is an FDA-registered facility (FEI 3004661493) located in SAN DIEGO, CA, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Nov 19, 2021: No Action Indicated (NAI)
- Inspections on record
- 6 (4 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Dec 17, 2025. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Aug 26, 2014
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 19, 2021 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Sep 13, 2018 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) Lack of Written MDR Procedures (21 CFR 803.17) |
| Dec 3, 2014 | Voluntary Action Indicated (VAI) | No citations listed |
| Aug 8, 2013 | Voluntary Action Indicated (VAI) | Quality plan (21 CFR 820.20(d)) Lack of or inadequate In-process acceptance procedures (21 CFR 820.80(c)) Quality Audit/Reaudit - conducted (21 CFR 820.22) Supplier oversight (21 CFR 820.50(a)(2)) and 1 more |
| Aug 20, 2010 | Voluntary Action Indicated (VAI) | Design control - no procedures (21 CFR 820.30(a)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) Lack of procedures, or not maintained (21 CFR 820.40) Environmental control Lack of or inadequate procedures (21 CFR 820.70(c)) |
| Jan 29, 2009 | No Action Indicated (NAI) | No citations listed |
Warning letters
- Dec 17, 2025: CGMP/QSR/Medical Devices/Adulterated. Read the letter on FDA.gov.
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 26, 2014 | Class II | Z-1076-2015 | The SurgiWrap Bioresorbable Adhesion Barrier Film. SurgiWrap 100x130x0.05mm
(01)… The expiration date of the bar code on the envelope label did not match the expiration date next to the hourglass. The printed expiration… |
FDA record score
32 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
6 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Warning letter, no close-out letter listed | Dec 17, 2025 | 30 x 1 | 30 |
| Inspection classified VAI | Sep 13, 2018 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Dec 3, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1076-2015, this FEI as the recalling firm | Aug 26, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 8, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 20, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |