INNO Holdings Inc. FDA Inspection Record

FEI 3004719693 · Inspections, Form 483s, warning letters and import refusals

INNO Holdings Inc. is an FDA-registered facility (FEI 3004719693) located in Millcreek, UT, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Aug 20, 2019: No Action Indicated (NAI)
Inspections on record
5 (2 VAI, 3 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
8, latest May 22, 2026

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Aug 20, 2019No Action Indicated (NAI)No citations listed
Jun 22, 2016No Action Indicated (NAI)No citations listed
May 22, 2014No Action Indicated (NAI)No citations listed
Apr 27, 2012Voluntary Action Indicated (VAI)Procedures to meet core CTGP (21 CFR 1271.180(a))
Storage temperatures recorded, maintained (21 CFR 1271.260(e))
Timely informing of consignees (21 CFR 1271.290(f))
Design validation - Risk analysis (21 CFR 820.30(g))
and 2 more
Jun 26, 2009Voluntary Action Indicated (VAI)Design control - no procedures (21 CFR 820.30(a))
Design input procedures - addressing conflicting requirement (21 CFR 820.30(c))
Acceptance criteria and essential outputs (21 CFR 820.30(d))
Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
and 8 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 22, 2026Class IIZ-2859-2026Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System
Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion…
Aug 14, 2024Class IIZ-3163-2024TxHA PEEK IBF/TxTiHA IBF System REF BTA-151 i. TxHA The Innovasis TxHA PEEK IBF…
Due to interfacing issues between the implant trial and inserter of the lumbar intervertebral fusion device.
Apr 7, 2020Class IIZ-2062-2020Oryx Cervical Screw Caddy, Model Number BTA-172, as part of the Oryx Cervical Plate System…
Cervical screw caddies have incorrect text markings for the 4.5mm Fixed and Variable rescue screws. The caddy location labeled as "Fixed"…
Jan 8, 2018Class IIZ-0644-2018Excella II Polyaxial Pedicle Screw used as a pedicle screw system in skeletally mature…
Nine lots of 8.5 mm Polyaxial Pedicle Screws are at risk of breakage.
May 5, 2017Class IIZ-2153-2017Opteryx Variable Rescue Screw, Part #CP2214, UDI: *+M711CP22140/$$801716012*. Packaged in…
Opteryx Variable Rescue Screws are engraved with incorrect (larger) size (16 instead of 14).
Dec 7, 2015Class IIZ-0662-2016Excella II Rod Cat. No.: SR 55100 5.5 x 100 mm LOT 1502 and Excella II Rod Cat. No.: SR…
Innovasis recalled two lots of Excella II titanium rods which were packaged using each other's labels.
Jul 7, 2014Class IIZ-2067-2014Excella II Standard Pedicle Screw, Catalog No. E2S47525 and E2S47535, 4.75 mm x 25 mm. For…
Innovasis is recalling the Excella II Standard Pedicle Screw due to mislabeling of the size and catalog number.
Jan 9, 2014Class IIIB-0802-14Bone
Human tissue allografts were distributed with expired packaging.

FDA record score

9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

9 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2859-2026, this FEI as the recalling firmMay 22, 20264 x 14
Class II recall Z-3163-2024, this FEI as the recalling firmAug 14, 20244 x 0.62.4
Class II recall Z-2062-2020, this FEI as the recalling firmApr 7, 20204 x 0.251
Class II recall Z-0644-2018, this FEI as the recalling firmJan 8, 20184 x 0.251
Class II recall Z-2153-2017, this FEI as the recalling firmMay 5, 20174 x 0.251
Class II recall Z-0662-2016, this FEI as the recalling firmDec 7, 201510 years or older, so it adds no points0
Class II recall Z-2067-2014, this FEI as the recalling firmJul 7, 201410 years or older, so it adds no points0
Inspection classified VAIApr 27, 201210 years or older, so it adds no points0
Inspection classified VAIJun 26, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing INNO Holdings Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check INNO Holdings Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Innovasis, Inc
Location
Millcreek, UT (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.