Lockheed Martin Gyrocam Systems, Inc. FDA Inspection Record
Lockheed Martin Gyrocam Systems, Inc. is an FDA-inspected facility (FEI 3004882148) located in Sarasota, FL, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Apr 8, 2011: No Action Indicated (NAI)
- Inspections on record
- 2 (1 OAI, 1 NAI)
- Last OAI inspection
- Jan 29, 2010
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Jan 21, 2010
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 8, 2011 | No Action Indicated (NAI) | No citations listed |
| Jan 29, 2010 | Official Action Indicated (OAI) | Failure to submit (21 CFR 1002.13) Certification tests inadequate (21 CFR 1040.10(e)(1)) Safety interlock not provided (21 CFR 1040.10(f)(2)(i)) Remote interlock connector (21 CFR 1040.10(f)(3)) and 3 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jan 21, 2010 | Class II | Z-0654-2013 | Lockheed Martin Gyrocam Systems ( LMGS) LMGS 15" camera systems. Laser pointer and… These camera systems were manufactured by LMGS and found to fail to comply with the FDA laser performance standard during an FDA… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
2 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified OAI | Jan 29, 2010 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0654-2013, this FEI as the recalling firm | Jan 21, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 27, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 27, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 27, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 27, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Lockheed Martin Gyrocam Systems, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Lockheed Martin Gyrocam Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Lockheed Martin Gyrocam Systems, Inc. → Check this supplier
Facility details
- Company
- Lockheed Martin Gyrocam Systems, Inc.
- Location
- Sarasota, FL (US)
- Source
- inspections