Covidien Llc FDA Inspection Record

FEI 3004904811 · Inspections, Form 483s, warning letters and import refusals

Covidien Llc is an FDA-registered facility (FEI 3004904811) located in Santa Clara, CA, US. RegDataLab tracks 9 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Sep 20, 2016: No Action Indicated (NAI)
Inspections on record
9 (6 VAI, 3 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
5, latest May 13, 2026

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 20, 2016No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))
Equipment Installation, Placement, Specified Requirements (21 CFR 820.70(g))
Jun 3, 2015No Action Indicated (NAI)No citations listed
May 16, 2013Voluntary Action Indicated (VAI)Investigation of device failures (21 CFR 820.198(c))
May 17, 2012Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Procedures not adequately established or maintained (21 CFR 820.40)
Environmental control Lack of or inadequate procedures (21 CFR 820.70(c))
Dec 5, 2011Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a))
Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))
Sep 2, 2010No Action Indicated (NAI)No citations listed
Jun 15, 2010Voluntary Action Indicated (VAI)Environmental control Lack of or inadequate procedures (21 CFR 820.70(c))
Report of Death or Serious Injury (21 CFR 803.50(a)(1))
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Conducting an Investigation (21 CFR 803.50(b)(3))
Jan 25, 2010Voluntary Action Indicated (VAI)Design validation unresolved discrepancies (21 CFR 820.30(g))
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Feb 23, 2009Voluntary Action Indicated (VAI)Training - Lack of or inadequate procedures (21 CFR 820.25(b))
Training (21 CFR 820.25(b))
Design input - documentation (21 CFR 820.30(c))
Documentation (21 CFR 820.75(a))
and 5 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 13, 2026Class IIZ-2950-2026EndoFLIP System that includes: ENDOFLIP 1 SYSTEM EF-100 05391530810050 10884521809307…
The system's 24-hour usage limit results in catheters incorrectly flagged as expired, making them unavailable for clinical procedures.
Apr 25, 2024Class IIZ-2038-2024EsoFLIP, 30mm BALLOON DILATION CATHETER, REF ES-330, Rx Only
Due to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements…
Jun 2, 2023Class IIZ-2282-2023Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell…
Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which…
Jun 6, 2016Class IIZ-2122-2016Covidien Cytosponge Cell Collection Kit. Contains Cytosponge Cell Collection Device… and 1 more product
Medtronic is recalling all lots of the Covidien Cytosponge Cell Collection device after two reports of the device detaching from the removal…
Aug 3, 2012Class IIZ-2379-2012HALO Flex Energy Generators Manufactured by B¿RRX Medical (now Covidien) 540 Oakmead…
Potential failure of HALO Energy Generator to enter "stand by" mode when initially powered on; and therefore can not be used for patient…

FDA record score

9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

11 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2950-2026, this FEI as the recalling firmMay 13, 20264 x 14
Class II recall Z-2038-2024, this FEI as the recalling firmApr 25, 20244 x 0.62.4
Class II recall Z-2282-2023, this FEI as the recalling firmJun 2, 20234 x 0.62.4
Class II recall Z-2122-2016, this FEI as the recalling firmJun 6, 201610 years or older, so it adds no points0
Inspection classified VAIMay 16, 201310 years or older, so it adds no points0
Class II recall Z-2379-2012, this FEI as the recalling firmAug 3, 201210 years or older, so it adds no points0
Inspection classified VAIMay 17, 201210 years or older, so it adds no points0
Inspection classified VAIDec 5, 201110 years or older, so it adds no points0
Inspection classified VAIJun 15, 201010 years or older, so it adds no points0
Inspection classified VAIJan 25, 201010 years or older, so it adds no points0
Inspection classified VAIFeb 23, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Covidien Llc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Covidien Llc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Covidien, LLC
Location
Santa Clara, CA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.