Covidien Llc FDA Inspection Record
Covidien Llc is an FDA-registered facility (FEI 3004904811) located in Santa Clara, CA, US. RegDataLab tracks 9 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Sep 20, 2016: No Action Indicated (NAI)
- Inspections on record
- 9 (6 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest May 13, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 20, 2016 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) Equipment Installation, Placement, Specified Requirements (21 CFR 820.70(g)) |
| Jun 3, 2015 | No Action Indicated (NAI) | No citations listed |
| May 16, 2013 | Voluntary Action Indicated (VAI) | Investigation of device failures (21 CFR 820.198(c)) |
| May 17, 2012 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Procedures not adequately established or maintained (21 CFR 820.40) Environmental control Lack of or inadequate procedures (21 CFR 820.70(c)) |
| Dec 5, 2011 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) |
| Sep 2, 2010 | No Action Indicated (NAI) | No citations listed |
| Jun 15, 2010 | Voluntary Action Indicated (VAI) | Environmental control Lack of or inadequate procedures (21 CFR 820.70(c)) Report of Death or Serious Injury (21 CFR 803.50(a)(1)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) Conducting an Investigation (21 CFR 803.50(b)(3)) |
| Jan 25, 2010 | Voluntary Action Indicated (VAI) | Design validation unresolved discrepancies (21 CFR 820.30(g)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Feb 23, 2009 | Voluntary Action Indicated (VAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Training (21 CFR 820.25(b)) Design input - documentation (21 CFR 820.30(c)) Documentation (21 CFR 820.75(a)) and 5 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 13, 2026 | Class II | Z-2950-2026 | EndoFLIP System that includes:
ENDOFLIP 1 SYSTEM EF-100
05391530810050
10884521809307… The system's 24-hour usage limit results in catheters incorrectly flagged as expired, making them unavailable for clinical procedures. |
| Apr 25, 2024 | Class II | Z-2038-2024 | EsoFLIP, 30mm BALLOON DILATION CATHETER, REF ES-330, Rx Only Due to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements… |
| Jun 2, 2023 | Class II | Z-2282-2023 | Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell… Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which… |
| Jun 6, 2016 | Class II | Z-2122-2016 | Covidien Cytosponge Cell Collection Kit.
Contains Cytosponge Cell Collection Device… and 1 more product Medtronic is recalling all lots of the Covidien Cytosponge Cell Collection device after two reports of the device detaching from the removal… |
| Aug 3, 2012 | Class II | Z-2379-2012 | HALO Flex Energy Generators
Manufactured by B¿RRX Medical (now Covidien)
540 Oakmead… Potential failure of HALO Energy Generator to enter "stand by" mode when initially powered on; and therefore can not be used for patient… |
FDA record score
9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
11 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2950-2026, this FEI as the recalling firm | May 13, 2026 | 4 x 1 | 4 |
| Class II recall Z-2038-2024, this FEI as the recalling firm | Apr 25, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-2282-2023, this FEI as the recalling firm | Jun 2, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-2122-2016, this FEI as the recalling firm | Jun 6, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 16, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2379-2012, this FEI as the recalling firm | Aug 3, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 17, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Dec 5, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 15, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 25, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Feb 23, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Covidien Llc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Covidien Llc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Covidien, LLC
- Location
- Santa Clara, CA (US)
- Source
- device