Samsung HME America, Inc. FDA Inspection Record
Samsung HME America, Inc. is an FDA-registered facility (FEI 3004938766) located in Danvers, MA, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Aug 27, 2024: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (4 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 13, latest Oct 7, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 27, 2024 | Voluntary Action Indicated (VAI) | Failure to submit (21 CFR 1002.13) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Jan 14, 2020 | Voluntary Action Indicated (VAI) | Failure to submit (21 CFR 1002.13) Failure to submit, distinct marking (21 CFR 1002.10) Results of tests (21 CFR 1002.30(a)(2)) |
| Jun 15, 2017 | Voluntary Action Indicated (VAI) | Failure to submit (21 CFR 1002.13) Failure to report (21 CFR 1002.20(a)) Lack of required elements (21 CFR 1002.13) |
| Mar 13, 2014 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) |
| Jan 27, 2012 | No Action Indicated (NAI) | Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 7, 2025 | Class II | Z-0480-2026 | GM85 Digital Mobile X-ray imaging System;
Model Number: GM85; The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue. |
| Jul 8, 2024 | Class II | Z-2715-2024 | GM60A. Digital Diagnostic Mobile X-Ray System. Potential for the anti-fall system of the device arm to fail. |
| Dec 28, 2023 | Class II | Z-1056-2024 | Samsung Digital Diagnostic Mobile X-ray System, Model GM85. Mobile x-ray systems moving arm frame welding issue which presents the potential of bodily harm due to fall of the arm. |
| Jun 6, 2023 | Class II | Z-2068-2023 | OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be… 1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in… |
| Mar 8, 2023 | Class II | Z-1414-2023 | Samsung GM85 Digital X-ray Imaging System-A Digital Diagnostic Mobile X-ray System Issue related to the operation of the arm latch within the column that supports the arm, potential bodily harm due to a fall of the arm |
| Jul 15, 2020 | Class II | Z-2772-2020 | OmniTom Computed tomography x-ray system, Model NL5000, all NL5000 systems with software… There are potential issues related to the operational state of the scanner, specifically for transport and inadvertent system shutdowns… |
| Jul 25, 2019 | Class II | Z-2275-2019 | HS70A Diagnostic Ultrasound System Version 2.01.00, 2.01.01, 2.01.02, 2.01.03, 2.01.04… and 1 more product There is a potential for probes overheating when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only… |
| May 6, 2019 | Class II | Z-1691-2019 | Samsung GM85 Mobile Digital X-ray Imaging System
Product Usage:
The GM85 Digital… Potential issue related to the operation of the safety latch within the column that supports the arm due to a possibility of potential… |
| Jan 12, 2018 | Class II | Z-1391-2018 | Solid state x-ray imager (flat panel/digital imager)
Solid state x-ray imager (flat… While transporting the portable machine with the retrofit equipment in a backwards motion, there is potential for foot injury from the… |
| Jul 28, 2017 | Class II | V-0096-2018 | PT Electrolyte Test 3V Samsung has recently determined that the polymer used in the reagent was over oxidized and has caused issues with stability for the Sodium… |
FDA record score
27 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
17 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0480-2026, this FEI as the recalling firm | Oct 7, 2025 | 4 x 1 | 4 |
| Inspection classified VAI | Aug 27, 2024 | 6 x 0.6 | 3.6 |
| Class II recall Z-2715-2024, this FEI as the recalling firm | Jul 8, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1056-2024, this FEI as the recalling firm | Dec 28, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-2068-2023, this FEI as the recalling firm | Jun 6, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1414-2023, this FEI as the recalling firm | Mar 8, 2023 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Jan 14, 2020 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Jun 15, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-2772-2020, this FEI as the recalling firm | Jul 15, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2275-2019, this FEI as the recalling firm | Jul 25, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1691-2019, this FEI as the recalling firm | May 6, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1391-2018, this FEI as the recalling firm | Jan 12, 2018 | 4 x 0.25 | 1 |
| Class II recall V-0096-2018, this FEI as the recalling firm | Jul 28, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0726-2018, this FEI as the recalling firm | Jun 22, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-3036-2017, this FEI as the recalling firm | May 24, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0263-2016, this FEI as the recalling firm | Sep 14, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 13, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Samsung HME America, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Samsung HME America, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- NeuroLogica Corporation
- Location
- Danvers, MA (US)
- Source
- device