Brasseler U.S.A. Medical, LLC FDA Inspection Record
Brasseler U.S.A. Medical, LLC is an FDA-registered facility (FEI 3004949275) located in Savannah, GA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- May 21, 2019: No Action Indicated (NAI)
- Inspections on record
- 3 (3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Jan 28, 2022
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| May 21, 2019 | No Action Indicated (NAI) | No citations listed |
| Sep 10, 2015 | No Action Indicated (NAI) | No citations listed |
| Jan 11, 2012 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jan 28, 2022 | Class II | Z-0707-2022 | Brasseler USA, K-WIRE, Double Trocar Smooth, REF KM172-26-35S One lot of product was distributed in unsealed packaging |
| Dec 11, 2019 | Class II | Z-0884-2020 | Brasseler USA KM169-39-76 Steinmann Pin Single Diamond Smooth, Diameter 2.8 mm, Length 228… 103 packaged and labeled tubes (103 product tubes with 6 pins per tube) of KM168-39-76 Single Diamond Smooth Steinmann Pins were labeled… |
| Oct 18, 2019 | Class II | Z-0591-2020 | BRASSELER KM166-310-35 Twist Drill Quick Connect Diameter 3.5 mm, Total Length 112mm… One lot of 3.5 mm twist drills were incorrectly laser etched as 2.5 mm the shaft of the device. |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
3 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0707-2022, this FEI as the recalling firm | Jan 28, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0884-2020, this FEI as the recalling firm | Dec 11, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0591-2020, this FEI as the recalling firm | Oct 18, 2019 | 4 x 0.25 | 1 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Brasseler U.S.A. Medical, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.