Med-Mizer, Inc FDA Inspection Record

FEI 3004976058 · Inspections, Form 483s, warning letters and import refusals

Med-Mizer, Inc is an FDA-registered facility (FEI 3004976058) located in BATESVILLE, IN, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Jul 31, 2026: No Action Indicated (NAI)
Inspections on record
5 (2 OAI, 3 NAI)
Last OAI inspection
Jul 6, 2015
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jul 22, 2014. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
3, latest Feb 3, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jul 31, 2026No Action Indicated (NAI)Risk-based approach not applied to QMS processes (ISO 13485:2016 Clause 4.1.2)
Arrangements for control of contaminated product not planned and documented (ISO 13485:2016 Clause 6.4.2)
Processes needed for product realization not planned and developed (ISO 13485:2016 Clause 7.1)
Production and service provision not adequate (ISO 13485:2016 Clause 7.5.1)
and 3 more
Aug 6, 2021No Action Indicated (NAI)Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Lack of or inadequate In-process acceptance procedures (21 CFR 820.80(c))
Lack of or inadequate DHR procedures (21 CFR 820.184)
and 4 more
Mar 28, 2018No Action Indicated (NAI)Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Design validation - simulated testing (21 CFR 820.30(g))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
and 1 more
Jul 6, 2015Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design input - Lack of or inadequate procedures (21 CFR 820.30(c))
Design output - Lack of or inadequate procedures (21 CFR 820.30(d))
Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
and 19 more
Mar 7, 2014Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Procedures not adequately established or maintained (21 CFR 820.40)
Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Training - Lack of or inadequate procedures (21 CFR 820.25(b))
and 13 more

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Feb 3, 2023Class IIZ-1204-2023Med-Riser MR600 Patient Lift and 1 more product
Risk of boom pivot failing due to the nut securing the boom becoming loose and the bolt to be unsecure to the flange on the side.
Jul 25, 2018Class IIZ-3183-2018PR1000 Pivot Rehab Bed AC powered adjustable bed.
Under certain conditions, when pivoting bed to the extreme left there are situations where the wire/cables may become pinched. A failure…
Apr 18, 2014Class IIZ-1541-2014RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2… and 2 more products
All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross…

FDA record score

3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 27, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

6 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1204-2023, this FEI as the recalling firmFeb 3, 20234 x 0.62.4
Class II recall Z-3183-2018, this FEI as the recalling firmJul 25, 20184 x 0.251
Inspection classified OAIJul 6, 201510 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJul 22, 201410 years or older, so it adds no points0
Class II recall Z-1541-2014, this FEI as the recalling firmApr 18, 201410 years or older, so it adds no points0
Inspection classified OAIMar 7, 201410 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 27, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 27, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 27, 2026Newest record Nov 19, 2026Current
Warning letter textNov 27, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 27, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Med-Mizer, Inc as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Med-Mizer, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Med-Mizer, Inc → Check this supplier

Facility details

FDA Legal Name
MED-MIZER, INC
Company
Med-Mizer, Inc.
Location
BATESVILLE, IN (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.