Xtant Medical Holdings, Inc. FDA Inspection Record

FEI 3005031160 · Inspections, Form 483s, warning letters and import refusals

Xtant Medical Holdings, Inc. is an FDA-registered facility (FEI 3005031160) located in Belgrade, MT, US. RegDataLab tracks 14 FDA inspections and 2 compliance actions at this facility.

What FDA found at this site

Latest inspection
Aug 13, 2021: No Action Indicated (NAI)
Inspections on record
14 (2 OAI, 2 VAI, 10 NAI)
Last OAI inspection
Jul 27, 2012
Warning letters
2 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jan 28, 2013. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
12, latest Mar 9, 2026

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Showing the 10 most recent of 14 inspections. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Aug 13, 2021No Action Indicated (NAI)No citations listed
Aug 25, 2017No Action Indicated (NAI)No citations listed
Aug 21, 2017No Action Indicated (NAI)No citations listed
May 18, 2016No Action Indicated (NAI)No citations listed
May 1, 2014No Action Indicated (NAI)No citations listed
Aug 2, 2013Voluntary Action Indicated (VAI)Procedures to meet core CTGP (21 CFR 1271.180(a))
Procedures for cleaning, sanitation (21 CFR 1271.190(d)(1))
Jul 12, 2013No Action Indicated (NAI)Lack of or inadequate procedures for handling (21 CFR 820.140)
Jul 11, 2013No Action Indicated (NAI)No citations listed
Jul 27, 2012Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate final acceptance procedures (21 CFR 820.80(d))
Lack of or inadequate procedures (21 CFR 820.100(a))
Not approved or obsolete document retrieval (21 CFR 820.40(a))
and 2 more
Jul 26, 2012Official Action Indicated (OAI)Procedures for all other requirements (21 CFR 1271.47(a))
Ensuring appropriate core requirements followed (21 CFR 1271.160(b)(1))
Procedures to meet core CTGP (21 CFR 1271.180(a))
Procedures for cleaning, sanitation (21 CFR 1271.190(d)(1))
and 8 more

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Showing the 10 most recent of 12 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Mar 9, 2026Class IIZ-1897-2026nanOss 3D Advanced Bone Graft Substitute nanOss 3D Plus Advanced Bone Graft…
Products were incorrectly labeled with a 5-year shelf life. The approved and validated shelf life is four years.
Feb 18, 2026Class IIZ-1773-2026XTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.
Units from one lot were mislabeled with an incorrect part number, identifying the device as a shorter screw length than it physically is.
Jan 3, 2025Class IIZ-1177-2025Irix-A Integrated Lumbar Fusion System, Model Number X080-423011-08PC (Spacer Assy, 42 X…
Certain implants in the Irix-A Lumbar Fusion System were distributed while labeled with an incorrect expiration date.
Aug 21, 2024Class IIZ-0125-2025Cortera Spinal Fixation System Screw, Poly, Solid, 6.5mm x 45mm, REF: 150100-65045
Spinal fixation system implant screws marked "DNI" are not intended to be implanted and may not have been manufactured in accordance with…
Nov 28, 2023Class IIZ-0846-2024Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5…
Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the…
Dec 29, 2021Class IIZ-0533-2022X060-0270, 8mm Titanium Insert, Axle Interspinous Fusion System.
The proximal opening of the inserts exhibit an out of specification condition which would prevent the crossbar plates from appropriately…
Sep 30, 2020Class IIIZ-0429-2021XTANT Medical, ## mm x ## mm x # mm x # ,Lordotic/Packaged InTice - C, Sterile R, Single…
Cervical Implants contain a label that incorrectly identifies the grade of titanium used.
May 12, 2020Class IIZ-2500-2020Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591
The strike plate fractured off from the forceps inserter devices with minimal impaction or force due to unsatisfactory weld.
May 17, 2016Class IIZ-0124-2017Calix P PEEK Lumbar System, PLIF Rasp and 3 more products
The trials and rasps used to prepare the surgical site for placement of the implant may become detached from the inserter assembly part and…
Apr 7, 2015Class IIZ-1823-2015Fortex" Pedicle Screw System Continuous Radius Rod 5.5 - 50mm, X-Spine Systems, Inc., 452…
The Continuous Radius Rod should measure 50mm in length but actually measures 45mm in length.

FDA record score

20 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 27, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

17 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1897-2026, this FEI as the recalling firmMar 9, 20264 x 14
Class II recall Z-1773-2026, this FEI as the recalling firmFeb 18, 20264 x 14
Class II recall Z-1177-2025, this FEI as the recalling firmJan 3, 20254 x 14
Class II recall Z-0125-2025, this FEI as the recalling firmAug 21, 20244 x 0.62.4
Class II recall Z-0846-2024, this FEI as the recalling firmNov 28, 20234 x 0.62.4
Class II recall Z-0533-2022, this FEI as the recalling firmDec 29, 20214 x 0.62.4
Class II recall Z-2500-2020, this FEI as the recalling firmMay 12, 20204 x 0.251
Class II recall Z-0124-2017, this FEI as the recalling firmMay 17, 201610 years or older, so it adds no points0
Class II recall Z-1823-2015, this FEI as the recalling firmApr 7, 201510 years or older, so it adds no points0
Class II recall Z-1512-2015, this FEI as the recalling firmMar 11, 201510 years or older, so it adds no points0
Class II recall Z-2136-2014, this FEI as the recalling firmJul 11, 201410 years or older, so it adds no points0
Inspection classified VAIAug 2, 201310 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJan 28, 201310 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJan 28, 201310 years or older, so it adds no points0
Inspection classified OAIJul 27, 201210 years or older, so it adds no points0
Inspection classified OAIJul 26, 201210 years or older, so it adds no points0
Inspection classified VAIAug 10, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 27, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 27, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 27, 2026Newest record Nov 19, 2026Current
Warning letter textNov 27, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 27, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Xtant Medical Holdings, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Xtant Medical Holdings, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Xtant Medical Holdings, Inc. → Check this supplier

Facility details

Company
XTANT Medical Holdings, Inc
Location
Belgrade, MT (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.