PENUMBRA, INC. FDA Inspection Record
PENUMBRA, INC. is an FDA-registered facility (FEI 3005168196) located in Alameda, CA, US. RegDataLab tracks 10 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jun 21, 2021: No Action Indicated (NAI)
- Inspections on record
- 10 (1 OAI, 4 VAI, 5 NAI)
- Last OAI inspection
- Aug 28, 2009
- Warning letters
- 1 listed by FDA for this FEI, latest Dec 31, 2009
- Recalls as the recalling firm
- 3, latest Jul 24, 2026
- Import refusals
- 2
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jun 21, 2021 | No Action Indicated (NAI) | No citations listed |
| Feb 13, 2017 | No Action Indicated (NAI) | No citations listed |
| Feb 23, 2016 | Voluntary Action Indicated (VAI) | No citations listed |
| Aug 4, 2015 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Jul 15, 2015 | Voluntary Action Indicated (VAI) | Sponsors' general responsibilities (21 CFR 812.40) |
| Apr 30, 2014 | No Action Indicated (NAI) | No citations listed |
| Feb 27, 2012 | No Action Indicated (NAI) | No citations listed |
| Aug 24, 2010 | Voluntary Action Indicated (VAI) | Report of Death or Serious Injury (21 CFR 803.50(a)(1)) |
| Aug 28, 2009 | Official Action Indicated (OAI) | Lack of or inadequate procedures for handling (21 CFR 820.140) Identification of actions needed (21 CFR 820.100(a)(3)) CAPA verification/validation of corrective/preventive action (21 CFR 820.100(a)(4)) Management responsibility (21 CFR 820.20) and 4 more |
| Feb 26, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 24, 2026 | Class II | Z-3331-2026 | Brand Name: Penumbra System (RED 62 Reperfusion Catheter)
Product Name: Penumbra System… Due to manufacturing issues reperfusion catheters may fracture. |
| Dec 15, 2020 | Class I | Z-0848-2021 | Penumbra JET 7 Reperfusion Catheter REF 5MAXJET7
The Penumbra JET 7 Reperfusion… and 3 more products The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast… |
| Jun 9, 2017 | Class I | Z-2702-2017 | Penumbra 3D Revascularization Device
It is indicated for used in the revascularization of… Penumbra has identified an issue in these four lots involving a raw material component of the delivery wire. This issue could result in… |
FDA record score
12 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-3331-2026, this FEI as the recalling firm | Jul 24, 2026 | 4 x 1 | 4 |
| Class I recall Z-0848-2021, this FEI as the recalling firm | Dec 15, 2020 | 15 x 0.25 | 3.8 |
| Class I recall Z-2702-2017, this FEI as the recalling firm | Jun 9, 2017 | 15 x 0.25 | 3.8 |
| Inspection classified VAI | Feb 23, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 4, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 15, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 24, 2010 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Dec 31, 2009 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Aug 28, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing PENUMBRA, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check PENUMBRA, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Penumbra Inc.
- Location
- Alameda, CA (US)
- Source