Otto Bock Healthcare Products GmbH FDA Inspection Record

FEI 3005190268 · Inspections, Form 483s, warning letters and import refusals

Otto Bock Healthcare Products GmbH is an FDA-registered facility (FEI 3005190268) located in Vienna Wien, AT. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Nov 21, 2019: Voluntary Action Indicated (VAI)
Inspections on record
3 (3 VAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
2, latest Sep 12, 2017

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Nov 21, 2019Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Software validation for automated processes (21 CFR 820.70(i))
Oct 19, 2012Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Lack of or inadequate DHR procedures (21 CFR 820.184)
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)
and 1 more
Oct 17, 2012Voluntary Action Indicated (VAI)Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Sep 12, 2017Class IIZ-0814-2018OTTO BOCK Kenevo knee joints Model 3C60 and 1 more product
Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has…
Aug 12, 2015Class IIZ-2438-2015Otto Bock Healthcare Products GmbH Genium X3 Model Number: 3B5-2, Manufacturer Otto Bock…
Otto Bock Healthcare Products GmbH is recalling the new Genium X3 knee joint prosthesis due to issues with the resistance-control of the…

FDA record score

3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 28, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

5 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAINov 21, 20196 x 0.251.5
Class II recall Z-0814-2018, this FEI as the recalling firmSep 12, 20174 x 0.251
Class II recall Z-2438-2015, this FEI as the recalling firmAug 12, 201510 years or older, so it adds no points0
Inspection classified VAIOct 19, 201210 years or older, so it adds no points0
Inspection classified VAIOct 17, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 28, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 28, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 28, 2026Newest record Nov 19, 2026Current
Warning letter textNov 28, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 28, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Otto Bock Healthcare Products GmbH as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Otto Bock Healthcare Products GmbH: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Otto Bock Healthcare Products GmbH → Check this supplier

Facility details

Company
Otto Bock Healthcare Products GmbH
Location
Vienna Wien (AT)
Source
device