Medtronic Vascular, Inc. FDA Inspection Record
Medtronic Vascular, Inc. is an FDA-inspected facility (FEI 3005364322) located in Santa Rosa, CA, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Sep 8, 2017: No Action Indicated (NAI)
- Inspections on record
- 5 (1 VAI, 4 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 7, latest Dec 21, 2021
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 8, 2017 | No Action Indicated (NAI) | No citations listed |
| Apr 19, 2016 | No Action Indicated (NAI) | No citations listed |
| Jan 21, 2016 | No Action Indicated (NAI) | No citations listed |
| Sep 4, 2013 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) Report of risk to health (21 CFR 806.10(a)(1)) Reporting Information Reasonably Known (21 CFR 803.50(b)(1)) |
| Jun 19, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Dec 21, 2021 | Class II | Z-0657-2022 | Endurant, Endurant II and Endurant IIs Stent Graft System:
Medtronic Endurant Stent… Stent Graft Systems built with specific batches of taper tip assemblies have the potential for the taper tip to detach from the delivery… |
| Oct 15, 2021 | Class II | Z-0290-2022 | Endurant Stent Graft System
REF/Description:
ENTF2525C70EE STENT GRAFT ENTF2525C70EE… and 2 more products During stent graft deployment, the spindle may detach from the distal end of the spindle hypotube |
| Jun 7, 2021 | Class II | Z-2135-2021 | Endurant II/Endurant IIs Stent Graft System:
Medtronic Endurant II Stent Graft System… and 1 more product During stent graft deployment, the radiopaque (RO) marker bond detached from the distal end of the graft cover. |
| Apr 28, 2021 | Class II | Z-1976-2021 | Endurant II/IIs Stent Graft System: STENT GRAFT ETBF3216C145E ENDUR II BIF. For treatment… Incorrect stentstop assembly (18fr) was used to build the delivery system component of the device. Correct stentstop assembly is 20fr. |
| Feb 12, 2021 | Class I | Z-1201-2021 | Valiant Navion Thoracic Stent Graft System. For endovascular repair of lesions in the… Due to the presence of type IIIb endoleaks, stent fractures, and stent ring enlargement. |
| Feb 27, 2017 | Class II | Z-1543-2017 | Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems
Model Numbers… Medtronic is initiating a voluntary Urgent Medical Device Recall for a subset of Endurant/ Endurant II Bifurcated Stent Graft Systems of… |
| Sep 19, 2012 | Class II | Z-0083-2014 | Valiant Thoracic Stent Graft with Captivia Delivery System,
Talent Thoracic Stent Graft… Firm received reports of difficulty in flushing the graft cover of the delivery system during device preparation as required by IFU. |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-1201-2021, this FEI as the recalling firm | Feb 12, 2021 | 15 x 0.25 | 3.8 |
| Class II recall Z-0657-2022, this FEI as the recalling firm | Dec 21, 2021 | 4 x 0.6 | 2.4 |
| Class II recall Z-0290-2022, this FEI as the recalling firm | Oct 15, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2135-2021, this FEI as the recalling firm | Jun 7, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1976-2021, this FEI as the recalling firm | Apr 28, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1543-2017, this FEI as the recalling firm | Feb 27, 2017 | 4 x 0.25 | 1 |
| Inspection classified VAI | Sep 4, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0083-2014, this FEI as the recalling firm | Sep 19, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |