BIOELECTRONICS CORPORATION FDA Inspection Record

FEI 3005550216 · Inspections, Form 483s, warning letters and import refusals

BIOELECTRONICS CORPORATION is an FDA-registered facility (FEI 3005550216) located in Jefferson, MD, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Jul 10, 2026: Voluntary Action Indicated (VAI)
Inspections on record
4 (2 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
1 listed by FDA for this FEI, latest Jan 31, 2012

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jul 10, 2026Voluntary Action Indicated (VAI)Product realization risk management processes not documented (ISO 13485:2016 Clause 7.1)
Procedures to control design and development changes not documented (ISO 13485:2016 Clause 7.3.9)
Purchased product verification activities not established or implemented (ISO 13485:2016 Clause 7.4.3)
Production and service provision not adequate (ISO 13485:2016 Clause 7.5.1)
and 2 more
Jan 16, 2020No Action Indicated (NAI)No citations listed
Jul 14, 2015No Action Indicated (NAI)Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
Apr 28, 2009Voluntary Action Indicated (VAI)Sponsor securing investigator compliance (21 CFR 812.46(a))
Sponsor records re non-significant risk status inadequate (21 CFR 812.140(b)(4)(ii))
Sponsor records re QSR/GMP & non-significant risk studies (21 CFR 812.140(b)(4)(v))
Sponsor notification of IRB approval withdrawal (21 CFR 812.150(b)(2))
and 1 more

FDA record score

6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 24, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

3 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIJul 10, 20266 x 16
Warning letter on FDA's dashboard, which does not show close-outsJan 31, 201210 years or older, so it adds no points0
Inspection classified VAIApr 28, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 24, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 24, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 24, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 24, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing BIOELECTRONICS CORPORATION as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check BIOELECTRONICS CORPORATION yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
BioElectronics Corporation
Location
Jefferson, MD (US)
Source
device