Materialise USA, LLC. FDA Inspection Record
Materialise USA, LLC. is an FDA-registered facility (FEI 3005718816) located in Plymouth, MI, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Oct 6, 2021: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (3 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Jul 13, 2022
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 6, 2021 | Voluntary Action Indicated (VAI) | Design input - Lack of or inadequate procedures (21 CFR 820.30(c)) Design plans - Lack of or inadequate (21 CFR 820.30(b)) Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) |
| Aug 1, 2017 | No Action Indicated (NAI) | No citations listed |
| May 12, 2015 | Voluntary Action Indicated (VAI) | No citations listed |
| Dec 10, 2013 | No Action Indicated (NAI) | Documentation (21 CFR 820.75(a)) Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) Documentation - review in response to changes or deviations (21 CFR 820.75(c)) Records not kept (21 CFR 806.20(a)) |
| Apr 22, 2011 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) and 3 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 13, 2022 | Class II | Z-1569-2022 | ProPlan CMF Anatomical Model, Maxilla-Cranium, Clear
Identification number SD900.208 MU22-INA-DUQ case was labeled as MU22-INA-DUQ but was shipped with the Anatomical Model of case MU22-NOF-SAK |
| May 3, 2016 | Class II | Z-1941-2016 | Signature TKA GDE/MDL Set 04-05
Patient-specific surgical guide and instrument set. A patient specific case of Signature Personalized Patient Care System Knee Guides included a femur guide which does not match the patient… |
| Oct 16, 2014 | Class II | Z-0418-2015 | Zimmer Patient Specific Instrument PERSONA CR Pin Guides (Femur and Tibia) is intended to… The box of guides for a specific case arrived intact but contained two (2) femoral guides instead of one (1) femoral guide and one (1)… |
| Nov 26, 2013 | Class II | Z-0628-2014 | REF 42-411571 QTY 1 CATALOG CAUTION. SEE INSTRUCTIONS FOR USE
SIGNATURE
M PKA GDE/MDL… Shipped femur guide contained a non-conformance which could possibly impact fit of the drill insert. |
| Nov 8, 2013 | Class II | Z-0577-2014 | ProPlan CMF Guide(s) Model(s)
REF SD900.008 Description: Single Use Only… and 1 more product Patient Specific Guide contained the incorrect case report. The case report contains patient specific surgical plan information. |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1569-2022, this FEI as the recalling firm | Jul 13, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Oct 6, 2021 | 6 x 0.25 | 1.5 |
| Class II recall Z-1941-2016, this FEI as the recalling firm | May 3, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 12, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0418-2015, this FEI as the recalling firm | Oct 16, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0628-2014, this FEI as the recalling firm | Nov 26, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0577-2014, this FEI as the recalling firm | Nov 8, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 22, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Materialise USA, LLC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Materialise USA, LLC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Materialise USA LLC
- Location
- Plymouth, MI (US)
- Source
- device