COOPERVISION INC. FDA Inspection Record
COOPERVISION INC. is an FDA-registered facility (FEI 3005724763) located in W Henrietta, NY, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Apr 11, 2018: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (1 OAI, 1 VAI, 2 NAI)
- Last OAI inspection
- Oct 24, 2011
- Warning letters
- 1 listed by FDA for this FEI, latest Dec 5, 2011
- Form 483 and inspection documents
- 1 in FDA's reading room
- Recalls as the recalling firm
- 13, latest Sep 22, 2026
- Import refusals
- 10 or more
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 11, 2018 | Voluntary Action Indicated (VAI) | Validation of changes to automated process software (21 CFR 820.70(i)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) |
| May 29, 2012 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Sampling methods - Lack of or inadequate procedures (21 CFR 820.250(b)) Procedures for product review,disposition lack of/inadequate (21 CFR 820.90(b)(1)) Management review - Lack of or inadequate procedures (21 CFR 820.20(c)) and 1 more |
| Oct 24, 2011 | Official Action Indicated (OAI) | Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of or inadequate DHR procedures (21 CFR 820.184) Investigation of device failures (21 CFR 820.198(c)) and 1 more |
| Jul 21, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 22, 2026 | Class II | Z-0204-2027 | Generic Name: Softmed Breathables Advanced Toric (fanfilcon A) Soft (Hydrophilic) Contact… The affected products bear an incorrect statement on the carton packaging indicating that the product is "Indicated for Extended Wear"… |
| Aug 12, 2026 | Class II | Z-3248-2026 | Brand Name: Pearle Vision 1 day
Product Name: (somofilcon A) Contact Lens
Model/Catalog… A small number of lenses within the affected lot are of the incorrect power. |
| Aug 22, 2025 | Class II | Z-2606-2025 | MyDay Toric, stenfilcon A) Contact Lens. Model/Catalog Number: MyDay Toric.
54% water… and 2 more products One lot manufactured with an invalid sterilization cycle. |
| Feb 27, 2025 | Class II | Z-1455-2025 | Brand Names: MyDay Toric, LensCrafters One Day Premium Daily Disposable Contact Lenses For… a limited number of lots were manufactured with an incorrect cylinder power. |
| Aug 29, 2024 | Class II | Z-3310-2024 | BIOFINITY XR TORIC Lenses - Soft Contact lenses Indicated for the correction of ametropia… and 1 more product Manufactured with a misaligned axis resulting in lenses with the incorrect power, may experience poor visual acuity |
| Feb 14, 2023 | Class II | Z-1282-2023 | Clariti 1-day toric soft contact lenses, Power -1.00 -1.25/90 The lenses manufactured in the affected lot were made with a misaligned axis resulting in lenses with the incorrect power. |
| Jan 26, 2023 | Class II | Z-1197-2023 | Biofinity XR Toric 6-pack- Soft (hydrophilic) Contact Lens (extended wear) and 1 more product Manufactured with misaligned axis resulting in lenses with the incorrect power, users may experience poor visual acuity |
| Dec 27, 2022 | Class II | Z-1014-2023 | MyDay Toric (90-pack)
Reveal 1 Day Toric (90-pack)
Soft (hydrophilic) Contact Lens… Lenses may be of the incorrect power. A patient who uses an affected lens may experience poor visual acuity. |
| May 19, 2021 | Class II | Z-2143-2021 | CLARITI 1 DAY TORIC Power: -07.50 -0.75/060
Daily Disposable Contact Lens with UV blocker and 11 more products Misaligned placement of an axis mark can cause the patient to experience blurred vision |
| Jan 13, 2020 | Class II | Z-1227-2020 | Coopervision Vertex Toric /2 Clear Toric-
Power -2.50 -0.75 /90
Soft (hydrophilic)… and 1 more product Incorrect power of lenses than the expected |
Form 483s and inspection documents
FDA record score
29 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
15 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0204-2027, this FEI as the recalling firm | Sep 22, 2026 | 4 x 1 | 4 |
| Class II recall Z-3248-2026, this FEI as the recalling firm | Aug 12, 2026 | 4 x 1 | 4 |
| Class II recall Z-2606-2025, this FEI as the recalling firm | Aug 22, 2025 | 4 x 1 | 4 |
| Class II recall Z-1455-2025, this FEI as the recalling firm | Feb 27, 2025 | 4 x 1 | 4 |
| Class II recall Z-3310-2024, this FEI as the recalling firm | Aug 29, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1282-2023, this FEI as the recalling firm | Feb 14, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1197-2023, this FEI as the recalling firm | Jan 26, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1014-2023, this FEI as the recalling firm | Dec 27, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Apr 11, 2018 | 6 x 0.25 | 1.5 |
| Class II recall Z-2143-2021, this FEI as the recalling firm | May 19, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1227-2020, this FEI as the recalling firm | Jan 13, 2020 | 4 x 0.25 | 1 |
| 80 import refusals in 2013 | Dec 31, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0542-2014, this FEI as the recalling firm | Nov 20, 2013 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Dec 5, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Oct 24, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing COOPERVISION INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check COOPERVISION INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- CooperVision, Inc.
- Location
- W Henrietta, NY (US)
- Source