Zimmer Trabecular Metal Technology, Inc. FDA Inspection Record

FEI 3005751028 · Inspections, Form 483s, warning letters and import refusals

Zimmer Trabecular Metal Technology, Inc. is an FDA-registered facility (FEI 3005751028) located in Parsippany, NJ, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Sep 1, 2017: No Action Indicated (NAI)
Inspections on record
5 (1 VAI, 4 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
5, latest Feb 1, 2019

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 1, 2017No Action Indicated (NAI)No citations listed
May 11, 2015Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Supplier notification of changes (21 CFR 820.50(b))
Production and Process Change Procedures, lack of or Inad. (21 CFR 820.70(b))
Jun 6, 2013No Action Indicated (NAI)Lack of or inadequate procedures (21 CFR 820.100(a))
May 21, 2012No Action Indicated (NAI)No citations listed
Apr 2, 2009No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Feb 1, 2019Class IIZ-1005-2019NexGenTrabecular Metal Tibial Half Block Augment, Right Lateral/Left Medial, Size 4 5MM… and 1 more product
Potentially comingled, resulting in the product in the box potentially not matching the product on the label
Jul 7, 2016Class IIIZ-1892-2017The TM Ardis Interbody System implant is a convex, straight TLIF or PLIF device for…
This field action is being initiated following the firm's discovery of a procedural error in the approval of literature resulting in the…
Apr 15, 2016Class IIZ-1950-2016POROUS PATELLA 32MM X 10MM - 00587806532 POROUS PATELLA 35MM X 10MM… and 5 more products
Zimmer Biomet is initiating a recall of specific lots of foil pouched, sterile Trabecular Metal implants that were manufactured between…
Feb 4, 2015Class IIZ-1325-2015NexGen TRABECULAR METAL FEMORAL AUGMENT BLOCK DISTAL ONLY, 20 MM Augment, Size D, Sterile…
One unit of the 20 mm augments was mislabeled as 5 mm and it was subsequently distributed.
Dec 2, 2014Class IIZ-0851-2015NexGen Complete Knee Solution Monoblock Tibial Provisional / Drill Guide Tibial Size 5…
During cleaning, two separate units of the NexGen Tibial Base plates did not fit with the Tibial Base Plate handle.

FDA record score

1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 16, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

5 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1005-2019, this FEI as the recalling firmFeb 1, 20194 x 0.251
Class II recall Z-1950-2016, this FEI as the recalling firmApr 15, 201610 years or older, so it adds no points0
Inspection classified VAIMay 11, 201510 years or older, so it adds no points0
Class II recall Z-1325-2015, this FEI as the recalling firmFeb 4, 201510 years or older, so it adds no points0
Class II recall Z-0851-2015, this FEI as the recalling firmDec 2, 201410 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 16, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 16, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 16, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 16, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing Zimmer Trabecular Metal Technology, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Zimmer Trabecular Metal Technology, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Zimmer Trabecular Metal Technology, Inc.
Location
Parsippany, NJ (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.