THORNHILL RESEARCH INC. FDA Inspection Record
THORNHILL RESEARCH INC. is an FDA-registered facility (FEI 3005803793) located in TORONTO Ontario, CA. RegDataLab tracks 1 FDA inspection at this facility.
What FDA found at this site
- Latest inspection
- Nov 19, 2015: Voluntary Action Indicated (VAI)
- Inspections on record
- 1 (1 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Nov 19, 2015
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 19, 2015 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Equipment suitability & capability (21 CFR 820.72(a)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) Report of risk to health (21 CFR 806.10(a)(1)) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Nov 19, 2015 | Class I | Z-0586-2016 | MOVES System, an emergency ventilator with suction, oxygen concentrator and… and 1 more product Batteries found unresponsive after storage, not recognized by main device, and are perpetually enabled. |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
2 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-0586-2016, this FEI as the recalling firm | Nov 19, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Nov 19, 2015 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing THORNHILL RESEARCH INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check THORNHILL RESEARCH INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Thornhill Research Inc
- Location
- TORONTO Ontario (CA)
- Source
- device