Keystone Dental Inc FDA Inspection Record

FEI 3005990499 · Inspections, Form 483s, warning letters and import refusals

Keystone Dental Inc is an FDA-inspected facility (FEI 3005990499) located in Burlington, MA, US. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Jun 8, 2017: Voluntary Action Indicated (VAI)
Inspections on record
3 (2 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
12, latest Feb 22, 2019

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jun 8, 2017Voluntary Action Indicated (VAI)Records of complaint investigation (21 CFR 820.198(e))
Lack of System for Event Evaluations (21 CFR 803.17(a)(1))
Report of Death or Serious Injury (21 CFR 803.50(a)(1))
Oct 7, 2013No Action Indicated (NAI)No citations listed
Oct 12, 2012Voluntary Action Indicated (VAI)Quality audits - auditor independence (21 CFR 820.22)
Lack of or inadequate procedures (21 CFR 820.100(a))
Report of risk to health (21 CFR 806.10(a)(1))

Recalls

Showing the 10 most recent of 12 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Feb 22, 2019Class IIZ-1061-2019Paltop Premium Surgical Kit-containing the 1.25mm Hex Driver For surgical and restorative… and 3 more products
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage…
May 1, 2018Class IIZ-2162-2018Prima Plus ¿4.1 x 11.5mm Implant, Catalog Number 15737K Keystone Dental implants are…
The implants are mislabeled.
Mar 13, 2018Class IIZ-1485-2018Genesis Surgical Cassette Tapered Implants are intended for placement following…
The surgical cassette contained extra-large white grommets making the 3.8 mm and 6.5mm dental burrs wobble within the grommets in the…
Jan 24, 2018Class IIZ-1443-2018Stage-1 Tissue Punch, Disposable, 4.5mm; catalog number: ISS141K. Intended for use in…
The cutting edge of tissue punches may be potentially damaged with small dents and nicks.
Oct 16, 2017Class IIIZ-1179-2018STAGE-1¿ RDS COC Abutment Plastic Sleeve, Non-Lock, Multi-Unit; Catalog Number…
A locking plastic abutment sleeve, used by dental technicians in dental laboratories to aid in the creation of the final restorative crown…
Aug 25, 2017Class IIZ-0727-2018Stage-1¿ RDS Indexed Abutment for RDS Stage-1¿ Single Stage Implants Keystone s…
The recalled units are mislabeled with an incorrect expiration date. The expiration date printed on the label is 2202-06. The correct…
Jun 20, 2017Class IIZ-0608-2018Multi-Unit Abutment Angled, 17¿, 4mm Cuff 4.1/4.5mm TiLobe¿ Implants, Catalog Number…
Abutments are unable to accept the final abutment screw
Jun 16, 2017Class IIZ-0718-2018Max Implant System - Bone Tap for use with 7.0mm diameter Max Implant, Catalog Number…
An 8mm diameter bone tap was laser marked as 7mm diameter and was packed with a 7mm label.
Mar 9, 2017Class IIZ-1818-2017Keystone Dental Inc. PrimaConnex Straight Implant TC, WD Internal Connection 5.0 x…
Straight PrimaConnex Implants contained Tapered PrimaConnex Implants
Sep 15, 2016Class IIZ-0299-2017Keystone Dental PrimaConnex¿ Straight Implant TC, RD Internal Connection 4.0 x…
Small diameter (SD) cover screw was packaged with the regular diameter implant and do not match the implant diameter (RD)

FDA record score

10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 26, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

13 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIJun 8, 20176 x 0.251.5
Class II recall Z-1061-2019, this FEI as the recalling firmFeb 22, 20194 x 0.251
Class II recall Z-2162-2018, this FEI as the recalling firmMay 1, 20184 x 0.251
Class II recall Z-1485-2018, this FEI as the recalling firmMar 13, 20184 x 0.251
Class II recall Z-1443-2018, this FEI as the recalling firmJan 24, 20184 x 0.251
Class II recall Z-0727-2018, this FEI as the recalling firmAug 25, 20174 x 0.251
Class II recall Z-0608-2018, this FEI as the recalling firmJun 20, 20174 x 0.251
Class II recall Z-0718-2018, this FEI as the recalling firmJun 16, 20174 x 0.251
Class II recall Z-1818-2017, this FEI as the recalling firmMar 9, 20174 x 0.251
Class II recall Z-0299-2017, this FEI as the recalling firmSep 15, 201610 years or older, so it adds no points0
Class II recall Z-0973-2013, this FEI as the recalling firmMar 11, 201310 years or older, so it adds no points0
Inspection classified VAIOct 12, 201210 years or older, so it adds no points0
Class II recall Z-0923-2013, this FEI as the recalling firmJul 5, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 26, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 26, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 26, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 26, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Keystone Dental Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Keystone Dental Inc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Keystone Dental Inc
Location
Burlington, MA (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.