Keystone Dental Inc FDA Inspection Record
Keystone Dental Inc is an FDA-inspected facility (FEI 3005990499) located in Burlington, MA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jun 8, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 12, latest Feb 22, 2019
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jun 8, 2017 | Voluntary Action Indicated (VAI) | Records of complaint investigation (21 CFR 820.198(e)) Lack of System for Event Evaluations (21 CFR 803.17(a)(1)) Report of Death or Serious Injury (21 CFR 803.50(a)(1)) |
| Oct 7, 2013 | No Action Indicated (NAI) | No citations listed |
| Oct 12, 2012 | Voluntary Action Indicated (VAI) | Quality audits - auditor independence (21 CFR 820.22) Lack of or inadequate procedures (21 CFR 820.100(a)) Report of risk to health (21 CFR 806.10(a)(1)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 22, 2019 | Class II | Z-1061-2019 | Paltop Premium Surgical Kit-containing the 1.25mm Hex Driver
For surgical and restorative… and 3 more products Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage… |
| May 1, 2018 | Class II | Z-2162-2018 | Prima Plus ¿4.1 x 11.5mm Implant, Catalog Number 15737K
Keystone Dental implants are… The implants are mislabeled. |
| Mar 13, 2018 | Class II | Z-1485-2018 | Genesis Surgical Cassette Tapered
Implants are intended for placement following… The surgical cassette contained extra-large white grommets making the 3.8 mm and 6.5mm dental burrs wobble within the grommets in the… |
| Jan 24, 2018 | Class II | Z-1443-2018 | Stage-1 Tissue Punch, Disposable, 4.5mm; catalog number: ISS141K.
Intended for use in… The cutting edge of tissue punches may be potentially damaged with small dents and nicks. |
| Oct 16, 2017 | Class III | Z-1179-2018 | STAGE-1¿ RDS COC Abutment Plastic Sleeve, Non-Lock, Multi-Unit; Catalog Number… A locking plastic abutment sleeve, used by dental technicians in dental laboratories to aid in the creation of the final restorative crown… |
| Aug 25, 2017 | Class II | Z-0727-2018 | Stage-1¿ RDS Indexed Abutment for RDS Stage-1¿ Single Stage
Implants
Keystone s… The recalled units are mislabeled with an incorrect expiration date. The expiration date printed on the label is 2202-06. The correct… |
| Jun 20, 2017 | Class II | Z-0608-2018 | Multi-Unit Abutment Angled, 17¿, 4mm Cuff 4.1/4.5mm TiLobe¿ Implants, Catalog Number… Abutments are unable to accept the final abutment screw |
| Jun 16, 2017 | Class II | Z-0718-2018 | Max Implant System - Bone Tap for use with 7.0mm diameter Max Implant, Catalog Number… An 8mm diameter bone tap was laser marked as 7mm diameter and was packed with a 7mm label. |
| Mar 9, 2017 | Class II | Z-1818-2017 | Keystone Dental Inc. PrimaConnex Straight Implant TC, WD Internal Connection 5.0 x… Straight PrimaConnex Implants contained Tapered PrimaConnex Implants |
| Sep 15, 2016 | Class II | Z-0299-2017 | Keystone Dental PrimaConnex¿ Straight Implant TC, RD Internal Connection 4.0 x… Small diameter (SD) cover screw was packaged with the regular diameter implant and do not match the implant diameter (RD) |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
13 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jun 8, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-1061-2019, this FEI as the recalling firm | Feb 22, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2162-2018, this FEI as the recalling firm | May 1, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1485-2018, this FEI as the recalling firm | Mar 13, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1443-2018, this FEI as the recalling firm | Jan 24, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0727-2018, this FEI as the recalling firm | Aug 25, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0608-2018, this FEI as the recalling firm | Jun 20, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0718-2018, this FEI as the recalling firm | Jun 16, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1818-2017, this FEI as the recalling firm | Mar 9, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0299-2017, this FEI as the recalling firm | Sep 15, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0973-2013, this FEI as the recalling firm | Mar 11, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 12, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0923-2013, this FEI as the recalling firm | Jul 5, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Keystone Dental Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Keystone Dental Inc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Keystone Dental Inc
- Location
- Burlington, MA (US)
- Source
- inspections