CAPTIVA SPINE, Inc. FDA Inspection Record
CAPTIVA SPINE, Inc. is an FDA-registered facility (FEI 3006082533) located in Jupiter, FL, US. RegDataLab tracks 6 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Feb 13, 2025: No Action Indicated (NAI)
- Inspections on record
- 6 (2 VAI, 4 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Feb 25, 2015
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 13, 2025 | No Action Indicated (NAI) | No citations listed |
| Jan 14, 2025 | No Action Indicated (NAI) | No citations listed |
| May 14, 2019 | No Action Indicated (NAI) | No citations listed |
| Feb 17, 2017 | No Action Indicated (NAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Jan 16, 2013 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Mar 9, 2012 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Design input - documentation (21 CFR 820.30(c)) Documentation (21 CFR 820.75(a)) and 4 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 25, 2015 | Class II | Z-1480-2015 | Captiva's lumbar intervertebral fusion system Reamers (color code rings on the Reamers). It is possible for the titanium color-coded ring to detach from the Reamers. |
| Sep 20, 2012 | Class II | Z-0413-2013 | CapLOX II Final Torque Driver, Indicator is a Pedicle screw spinal system.
The CapLOX… Captiva Spine recalled their CapLOX II Final Torque Driver, Indicator due to the device prematurely breaking |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1480-2015, this FEI as the recalling firm | Feb 25, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 16, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0413-2013, this FEI as the recalling firm | Sep 20, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 9, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |