EVENT MEDICAL, LTD. FDA Inspection Record
EVENT MEDICAL, LTD. is an FDA-registered facility (FEI 3006135941) located in Lake Forest, CA, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- May 18, 2022: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (3 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Oct 13, 2015
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| May 18, 2022 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) |
| Aug 1, 2019 | No Action Indicated (NAI) | No citations listed |
| Sep 24, 2014 | Voluntary Action Indicated (VAI) | Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Sep 16, 2011 | No Action Indicated (NAI) | No citations listed |
| Oct 8, 2010 | Voluntary Action Indicated (VAI) | Design plans - Lack of or inadequate (21 CFR 820.30(b)) Design input - documentation (21 CFR 820.30(c)) Design validation - production units (21 CFR 820.30(g)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 13, 2015 | Class I | Z-0230-2016 | Inspiration Ventilator System, i Series and LS 12.1 LCD. The Inspiration 7i / 5i… A component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming. |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | May 18, 2022 | 6 x 0.6 | 3.6 |
| Class I recall Z-0230-2016, this FEI as the recalling firm | Oct 13, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Sep 24, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 8, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing EVENT MEDICAL, LTD. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.