EOS Imaging FDA Inspection Record
EOS Imaging is an FDA-registered facility (FEI 3006165260) located in PARIS, FR. RegDataLab tracks 1 FDA inspection at this facility.
What FDA found at this site
- Latest inspection
- Mar 15, 2012: Voluntary Action Indicated (VAI)
- Inspections on record
- 1 (1 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Dec 22, 2020
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 15, 2012 | Voluntary Action Indicated (VAI) | DHR content (21 CFR 820.184) Incoming acceptance records, documentation (21 CFR 820.80(b)) Failure to submit (21 CFR 1002.11) Filtration located in useful beam during exposure (21 CFR 1020.30(m)(1)) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Dec 22, 2020 | Class II | Z-1415-2021 | The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or… Inadequate images resizing and 2D measurement errors may occur when biplanar acquisition has been performed with patient orientation… |
| Aug 1, 2018 | Class II | Z-3188-2018 | sterEOS workstation 3D projections may be incorrect when the pair of images used is a secondary capture generated from a series containing more than two… |
| May 30, 2017 | Class II | Z-2626-2017 | ROTH MOBELI Grab Handles used with EOS System: MOBELI Stabi Vario Grab Handle, ROTH… ROTH MOBELI grab handles could detach and patient could fall if the EOS system operator does not follow instructions. Before each use, the… |
| Nov 29, 2013 | Class II | Z-0580-2014 | steroEOS Workstation with software versions: 1.3, 1.4 and 1.5 used in processing 3D X-ray… Error copying information to the clinical file and to the report when two 3D models are opened simultaneously on the sterEOS workstation |
| Nov 7, 2013 | Class II | Z-0529-2014 | EOS System X- ray beam
Digital radiography system used in general radiographic… EOS imaging discovered during production internal testing that the X-ray beam collimation set up by the operator at the user interface of… |
FDA record score
3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
6 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1415-2021, this FEI as the recalling firm | Dec 22, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-3188-2018, this FEI as the recalling firm | Aug 1, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2626-2017, this FEI as the recalling firm | May 30, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0580-2014, this FEI as the recalling firm | Nov 29, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0529-2014, this FEI as the recalling firm | Nov 7, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 15, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing EOS Imaging as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.