ORTHOPEDIATRICS, CORP FDA Inspection Record

FEI 3006460162 · Inspections, Form 483s, warning letters and import refusals

ORTHOPEDIATRICS, CORP is an FDA-registered facility (FEI 3006460162) located in Warsaw, IN, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Aug 18, 2023: Voluntary Action Indicated (VAI)
Inspections on record
5 (3 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
21, latest Sep 12, 2022
Import refusals
10 or more

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Aug 18, 2023Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Mar 23, 2017Voluntary Action Indicated (VAI)Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
Lack of or inadequate procedures (21 CFR 820.100(a))
Report of risk to health (21 CFR 806.10(a)(1))
and 2 more
Jun 19, 2014No Action Indicated (NAI)No citations listed
Mar 11, 2011Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Mar 6, 2009No Action Indicated (NAI)Lack of or inadequate procedures (21 CFR 820.100(a))

Recalls

Showing the 10 most recent of 21 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Sep 12, 2022Class IIZ-0277-2023Orthex Large Bone Shoulder Bolt, model no. AS-17
Bolt breakage on the head or threaded portion, which may cause loss of strut continuity with frame. This may cause loss of deformity…
Jun 29, 2022Class IIZ-1678-2022Orthex External Fixation Wire Sharp Bayonet 1.8 x 400 mm Model: K400-18BN
Manufacturing variance near transition of tip is increasing the likelihood the wire tip may fracture. The fractured tip may be retained in…
Jul 6, 2020Class IIZ-2742-2020Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System…
Product may have incorrect laser etched alignment lines, the non-conforming laser etch may have compromised strength, lead to unintended…
Jun 5, 2020Class IIZ-2538-2020OrthoPediatrics ACL Reconstruction System ShieldLoc 6mm implant component model number… and 1 more product
ShieldLoc implant component of the ACL Reconstruction Systemlots may be labeled with the incorrect size. Product labeled as 10-1008-4070 was…
Aug 19, 2019Class IIZ-0125-2020OrthoPediatrics Adolescent attachment bolt used with the Pediatric Nailing Platform (PNP)…
Interference can occur between the adolescent attachment bolt and the adolescent nail, cause a delay in surgical procedures
Aug 19, 2019Class IIZ-0218-2020Model No. 01-1501-0032 (Child Nail Attachment Bolt). Component of the Pediatric Nailing… and 1 more product
Binding can occur between the attachment bolt and intramedullary nail, which may lead to delays in surgical procedures.
Aug 3, 2017Class IIZ-0597-2018Response 5.5/6.0 Spine System 7.0mm x 40mm Pedicle Screw Product Usage: Posterior…
The 7mm diameter uniaxial pedicle screws have been color anodized with the same color utilized on 5 mm diameter screws.
Apr 24, 2017Class IIZ-2321-2017Response BandLoc 5.5/6.0 Tensioner Instrument The Tensioner Instrument (01-1102-1000)…
Over-tensioning of the band resulting in damage to the band causing it to rupture
Nov 18, 2016Class IIIZ-0879-2017OrthoPediatrics PediNail Intramedullary Nailing System 4.5mm X 48mm Cortical Screw
Labeled as a 48mm Cortical Screw but measuring at 50mm
Jul 27, 2016Class IIZ-2843-20163.5mm Bowed Locking Compression Femur Plates, 14 hole Orthopediatrics PediLoc Locking… and 1 more product
OrthoPediatrics Corp.is voluntarily recalling multiple lots of the 3.5mm Bowed Locking Compression Femur 14-hole Plate and 18-hole Plate due…

FDA record score

16 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

24 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIAug 18, 20236 x 0.63.6
Class II recall Z-0277-2023, this FEI as the recalling firmSep 12, 20224 x 0.62.4
Class II recall Z-1678-2022, this FEI as the recalling firmJun 29, 20224 x 0.62.4
Inspection classified VAIMar 23, 20176 x 0.251.5
Class II recall Z-2742-2020, this FEI as the recalling firmJul 6, 20204 x 0.251
Class II recall Z-2538-2020, this FEI as the recalling firmJun 5, 20204 x 0.251
Class II recall Z-0125-2020, this FEI as the recalling firmAug 19, 20194 x 0.251
Class II recall Z-0218-2020, this FEI as the recalling firmAug 19, 20194 x 0.251
Class II recall Z-0597-2018, this FEI as the recalling firmAug 3, 20174 x 0.251
Class II recall Z-2321-2017, this FEI as the recalling firmApr 24, 20174 x 0.251
13 import refusals in 2019Dec 31, 20192 x 0.250.5
Class II recall Z-2843-2016, this FEI as the recalling firmJul 27, 201610 years or older, so it adds no points0
Class II recall Z-0622-2016, this FEI as the recalling firmNov 10, 201510 years or older, so it adds no points0
Class II recall Z-2796-2015, this FEI as the recalling firmJul 28, 201510 years or older, so it adds no points0
Class II recall Z-2560-2014, this FEI as the recalling firmJul 25, 201410 years or older, so it adds no points0
Class II recall Z-0934-2014, this FEI as the recalling firmJan 8, 201410 years or older, so it adds no points0
Class II recall Z-0579-2014, this FEI as the recalling firmNov 7, 201310 years or older, so it adds no points0
Class II recall Z-0276-2014, this FEI as the recalling firmSep 26, 201310 years or older, so it adds no points0
Class II recall Z-1941-2013, this FEI as the recalling firmJul 24, 201310 years or older, so it adds no points0
Class II recall Z-1534-2013, this FEI as the recalling firmApr 19, 201310 years or older, so it adds no points0
Class II recall Z-0432-2013, this FEI as the recalling firmOct 18, 201210 years or older, so it adds no points0
Class II recall Z-0207-2013, this FEI as the recalling firmSep 13, 201210 years or older, so it adds no points0
Class II recall Z-2057-2012, this FEI as the recalling firmJun 14, 201210 years or older, so it adds no points0
Inspection classified VAIMar 11, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing ORTHOPEDIATRICS, CORP as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check ORTHOPEDIATRICS, CORP yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
OrthoPediatrics Corp
Location
Warsaw, IN (US)
Source
device