ORTHOPEDIATRICS, CORP FDA Inspection Record
ORTHOPEDIATRICS, CORP is an FDA-registered facility (FEI 3006460162) located in Warsaw, IN, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Aug 18, 2023: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (3 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 21, latest Sep 12, 2022
- Import refusals
- 10 or more
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 18, 2023 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) |
| Mar 23, 2017 | Voluntary Action Indicated (VAI) | Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) Report of risk to health (21 CFR 806.10(a)(1)) and 2 more |
| Jun 19, 2014 | No Action Indicated (NAI) | No citations listed |
| Mar 11, 2011 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) |
| Mar 6, 2009 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 12, 2022 | Class II | Z-0277-2023 | Orthex Large Bone Shoulder Bolt, model no. AS-17 Bolt breakage on the head or threaded portion, which may cause loss of strut continuity with frame. This may cause loss of deformity… |
| Jun 29, 2022 | Class II | Z-1678-2022 | Orthex External Fixation Wire Sharp Bayonet 1.8 x 400 mm
Model: K400-18BN Manufacturing variance near transition of tip is increasing the likelihood the wire tip may fracture. The fractured tip may be retained in… |
| Jul 6, 2020 | Class II | Z-2742-2020 | Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System… Product may have incorrect laser etched alignment lines, the non-conforming laser etch may have compromised strength, lead to unintended… |
| Jun 5, 2020 | Class II | Z-2538-2020 | OrthoPediatrics ACL Reconstruction System ShieldLoc 6mm implant component model number… and 1 more product ShieldLoc implant component of the ACL Reconstruction Systemlots may be labeled with the incorrect size. Product labeled as 10-1008-4070 was… |
| Aug 19, 2019 | Class II | Z-0125-2020 | OrthoPediatrics Adolescent attachment bolt used with the Pediatric Nailing Platform (PNP)… Interference can occur between the adolescent attachment bolt and the adolescent nail, cause a delay in surgical procedures |
| Aug 19, 2019 | Class II | Z-0218-2020 | Model No. 01-1501-0032 (Child Nail Attachment Bolt). Component of the Pediatric Nailing… and 1 more product Binding can occur between the attachment bolt and intramedullary nail, which may lead to delays in surgical procedures. |
| Aug 3, 2017 | Class II | Z-0597-2018 | Response 5.5/6.0 Spine System 7.0mm x 40mm
Pedicle Screw
Product Usage:
Posterior… The 7mm diameter uniaxial pedicle screws have been color anodized with the same color utilized on 5 mm diameter screws. |
| Apr 24, 2017 | Class II | Z-2321-2017 | Response BandLoc 5.5/6.0 Tensioner Instrument
The Tensioner Instrument (01-1102-1000)… Over-tensioning of the band resulting in damage to the band causing it to rupture |
| Nov 18, 2016 | Class III | Z-0879-2017 | OrthoPediatrics PediNail Intramedullary Nailing System
4.5mm X 48mm Cortical Screw Labeled as a 48mm Cortical Screw but measuring at 50mm |
| Jul 27, 2016 | Class II | Z-2843-2016 | 3.5mm Bowed Locking Compression Femur Plates, 14 hole
Orthopediatrics PediLoc Locking… and 1 more product OrthoPediatrics Corp.is voluntarily recalling multiple lots of the 3.5mm Bowed Locking Compression Femur 14-hole Plate and 18-hole Plate due… |
FDA record score
16 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
24 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Aug 18, 2023 | 6 x 0.6 | 3.6 |
| Class II recall Z-0277-2023, this FEI as the recalling firm | Sep 12, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1678-2022, this FEI as the recalling firm | Jun 29, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Mar 23, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-2742-2020, this FEI as the recalling firm | Jul 6, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2538-2020, this FEI as the recalling firm | Jun 5, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0125-2020, this FEI as the recalling firm | Aug 19, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0218-2020, this FEI as the recalling firm | Aug 19, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0597-2018, this FEI as the recalling firm | Aug 3, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2321-2017, this FEI as the recalling firm | Apr 24, 2017 | 4 x 0.25 | 1 |
| 13 import refusals in 2019 | Dec 31, 2019 | 2 x 0.25 | 0.5 |
| Class II recall Z-2843-2016, this FEI as the recalling firm | Jul 27, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0622-2016, this FEI as the recalling firm | Nov 10, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2796-2015, this FEI as the recalling firm | Jul 28, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2560-2014, this FEI as the recalling firm | Jul 25, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0934-2014, this FEI as the recalling firm | Jan 8, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0579-2014, this FEI as the recalling firm | Nov 7, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0276-2014, this FEI as the recalling firm | Sep 26, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1941-2013, this FEI as the recalling firm | Jul 24, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1534-2013, this FEI as the recalling firm | Apr 19, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0432-2013, this FEI as the recalling firm | Oct 18, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0207-2013, this FEI as the recalling firm | Sep 13, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2057-2012, this FEI as the recalling firm | Jun 14, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 11, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing ORTHOPEDIATRICS, CORP as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check ORTHOPEDIATRICS, CORP yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- OrthoPediatrics Corp
- Location
- Warsaw, IN (US)
- Source
- device