NORDICNEUROLAB AS FDA Inspection Record

FEI 3006738448 · Inspections, Form 483s, warning letters and import refusals

NORDICNEUROLAB AS is an FDA-registered facility (FEI 3006738448) located in BERGEN Hordaland, NO. RegDataLab tracks 1 FDA inspection at this facility.

What FDA found at this site

Latest inspection
Sep 13, 2012: No Action Indicated (NAI)
Inspections on record
1 (1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
8, latest Jun 16, 2022

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 13, 2012No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jun 16, 2022Class IIZ-1462-2022nordicBrainEX, software versions 2.21 through 2.3.10
When exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both the underlay and overlay originated from a…
Aug 20, 2018Class IIZ-3023-2018fMRI Hardware System VisualSystem allows video signals from the stimulus presentation PC…
Complaints of this version of the power supply unit (LPS) (1.0 and 1.1) being dragged into the magnet bore (MR Scanner).
Oct 28, 2017Class IIZ-2103-2018fMRI Hardware System - AudioSystem, Version 1.0 The AudioSystem allows auditory signals…
Report was of an interrupted MR scan after patient complained about acoustic distortion, and heat felt form the headphones.
Jul 3, 2015Class IIZ-2061-2018nordicICE v 2.3.14 nordicICE is an image processing software package to be used by…
The bug causes BOLD activation maps to be visualized as overlays without taking the coregistration into account. Any output created from…
Dec 15, 2014Class IIZ-2042-2018nordicTumorEx 1.0 nordicTumorEx is an image processing software package to be used by…
An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and…
Sep 17, 2014Class IIZ-2050-2018nordicBrainEx 2.0 provides analysis and visualization capabilities of dynamic MRI data of…
An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and…
Sep 12, 2014Class IIZ-2044-2018nordicICE 2.3.14 Image processing software package used by trained professionals…
An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and…
Dec 21, 2012Class IIZ-2045-2018nordicBrainEX Product Usage: nordicBrainEx is an image processing software package to…
There is an error related to the relative geometry between fiber tracts in a fiber group and images. The error occurs in the following…

FDA record score

4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 26, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

8 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1462-2022, this FEI as the recalling firmJun 16, 20224 x 0.62.4
Class II recall Z-3023-2018, this FEI as the recalling firmAug 20, 20184 x 0.251
Class II recall Z-2103-2018, this FEI as the recalling firmOct 28, 20174 x 0.251
Class II recall Z-2061-2018, this FEI as the recalling firmJul 3, 201510 years or older, so it adds no points0
Class II recall Z-2042-2018, this FEI as the recalling firmDec 15, 201410 years or older, so it adds no points0
Class II recall Z-2050-2018, this FEI as the recalling firmSep 17, 201410 years or older, so it adds no points0
Class II recall Z-2044-2018, this FEI as the recalling firmSep 12, 201410 years or older, so it adds no points0
Class II recall Z-2045-2018, this FEI as the recalling firmDec 21, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 26, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 26, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 26, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 26, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing NORDICNEUROLAB AS as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check NORDICNEUROLAB AS yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
NordicNeuroLab AS
Location
BERGEN Hordaland (NO)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.