CME America, LLC FDA Inspection Record

FEI 3006967710 · Inspections, Form 483s, warning letters and import refusals

CME America, LLC is an FDA-inspected facility (FEI 3006967710) located in Golden, CO, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Apr 23, 2018: No Action Indicated (NAI)
Inspections on record
5 (1 OAI, 2 VAI, 2 NAI)
Last OAI inspection
Feb 3, 2016
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
7, latest Jul 30, 2020

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Apr 23, 2018No Action Indicated (NAI)Design verification - output does not meet input requirement (21 CFR 820.30(f))
Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Lack of or inadequate procedures (21 CFR 820.100(a))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)
Aug 1, 2017Voluntary Action Indicated (VAI)Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
DMR - not or inadequately maintained (21 CFR 820.181)
Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1))
Lack of or inadequate process validation (21 CFR 820.75(a))
Feb 3, 2016Official Action Indicated (OAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Investigation of device failures (21 CFR 820.198(c))
Service reports (21 CFR 820.200(d))
Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1))
and 1 more
Aug 6, 2013Voluntary Action Indicated (VAI)Investigation of device failures (21 CFR 820.198(c))
DMR - not or inadequately maintained (21 CFR 820.181)
Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1))
Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))
Jul 23, 2010No Action Indicated (NAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Investigation of device failures (21 CFR 820.198(c))
Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 30, 2020Class IIZ-0525-2021CME America T-Syringe Pump -T34L - Product Usage: infusion of medications or fluids… and 5 more products
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being…
Jun 16, 2020Class IZ-2553-2020BodyGuard BodySet, REF: A120-160SAS and 36 more products
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion…
Feb 27, 2020Class IIZ-0305-2021CMEAmerica BodyGuard Wall Charger: Lot number 2019-0364, Mode/Catalogue number 151-143XL…
The wall charger may not properly charge the infusion pump battery.
Jan 14, 2020Class IIZ-1886-2020BodyGuard Microset For Use with the BodyGuard Infusion Pump *** Tubing: 110 in. (280 cm)…
The filter installed onto the tube set may have the incorrect pore size.
Jan 7, 2020Class IZ-1474-2020CME America BodyGuard Infusion Pump System, All models of BG 323, BG 121, BG 545, BG…
Infusion Pump Systems may have a delivery inaccuracy of up to 13%, which may result in 1) faster than expected drug delivery when infusing…
Sep 16, 2019Class IZ-1442-2020BodyGuard Microset REF A120-003XYVA, Sterile EO, Tubing: 150 in. (380 cm), Prime Vol…
When using a specific tubing set with infusion pump system, may result in under deliver of fluids.
Jan 16, 2015Class IIZ-1126-2015BodyGuard 323 pump, models 100-510PXSI, 100-516PXS, 100-517PXS, 100-518PXS, 100-603XSA…
CME America is recalling the BodyGuard and BodyGuard 323 Infusion pumps due to the potential for an over delivery.

FDA record score

16 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 24, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-2553-2020, this FEI as the recalling firmJun 16, 202015 x 0.253.8
Class I recall Z-1474-2020, this FEI as the recalling firmJan 7, 202015 x 0.253.8
Class I recall Z-1442-2020, this FEI as the recalling firmSep 16, 201915 x 0.253.8
Inspection classified VAIAug 1, 20176 x 0.251.5
Class II recall Z-0525-2021, this FEI as the recalling firmJul 30, 20204 x 0.251
Class II recall Z-0305-2021, this FEI as the recalling firmFeb 27, 20204 x 0.251
Class II recall Z-1886-2020, this FEI as the recalling firmJan 14, 20204 x 0.251
Inspection classified OAIFeb 3, 201610 years or older, so it adds no points0
Class II recall Z-1126-2015, this FEI as the recalling firmJan 16, 201510 years or older, so it adds no points0
Inspection classified VAIAug 6, 201310 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 24, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 24, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 24, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 24, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing CME America, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check CME America, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key