CME America, LLC FDA Inspection Record
CME America, LLC is an FDA-inspected facility (FEI 3006967710) located in Golden, CO, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Apr 23, 2018: No Action Indicated (NAI)
- Inspections on record
- 5 (1 OAI, 2 VAI, 2 NAI)
- Last OAI inspection
- Feb 3, 2016
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 7, latest Jul 30, 2020
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 23, 2018 | No Action Indicated (NAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Lack of or inadequate procedures (21 CFR 820.100(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Aug 1, 2017 | Voluntary Action Indicated (VAI) | Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) DMR - not or inadequately maintained (21 CFR 820.181) Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Feb 3, 2016 | Official Action Indicated (OAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Investigation of device failures (21 CFR 820.198(c)) Service reports (21 CFR 820.200(d)) Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1)) and 1 more |
| Aug 6, 2013 | Voluntary Action Indicated (VAI) | Investigation of device failures (21 CFR 820.198(c)) DMR - not or inadequately maintained (21 CFR 820.181) Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) |
| Jul 23, 2010 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 30, 2020 | Class II | Z-0525-2021 | CME America T-Syringe Pump -T34L - Product Usage: infusion of medications or fluids… and 5 more products Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being… |
| Jun 16, 2020 | Class I | Z-2553-2020 | BodyGuard BodySet, REF: A120-160SAS and 36 more products Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion… |
| Feb 27, 2020 | Class II | Z-0305-2021 | CMEAmerica BodyGuard Wall Charger: Lot number 2019-0364, Mode/Catalogue number 151-143XL… The wall charger may not properly charge the infusion pump battery. |
| Jan 14, 2020 | Class II | Z-1886-2020 | BodyGuard Microset For Use with the BodyGuard Infusion Pump *** Tubing: 110 in. (280 cm)… The filter installed onto the tube set may have the incorrect pore size. |
| Jan 7, 2020 | Class I | Z-1474-2020 | CME America BodyGuard Infusion Pump System, All models of BG 323, BG 121, BG 545, BG… Infusion Pump Systems may have a delivery inaccuracy of up to 13%, which may result in 1) faster than expected drug delivery when infusing… |
| Sep 16, 2019 | Class I | Z-1442-2020 | BodyGuard Microset REF A120-003XYVA, Sterile EO, Tubing: 150 in. (380 cm), Prime Vol… When using a specific tubing set with infusion pump system, may result in under deliver of fluids. |
| Jan 16, 2015 | Class II | Z-1126-2015 | BodyGuard 323 pump, models
100-510PXSI, 100-516PXS, 100-517PXS, 100-518PXS,
100-603XSA… CME America is recalling the BodyGuard and BodyGuard 323 Infusion pumps due to the potential for an over delivery. |
FDA record score
16 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
10 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-2553-2020, this FEI as the recalling firm | Jun 16, 2020 | 15 x 0.25 | 3.8 |
| Class I recall Z-1474-2020, this FEI as the recalling firm | Jan 7, 2020 | 15 x 0.25 | 3.8 |
| Class I recall Z-1442-2020, this FEI as the recalling firm | Sep 16, 2019 | 15 x 0.25 | 3.8 |
| Inspection classified VAI | Aug 1, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0525-2021, this FEI as the recalling firm | Jul 30, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0305-2021, this FEI as the recalling firm | Feb 27, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1886-2020, this FEI as the recalling firm | Jan 14, 2020 | 4 x 0.25 | 1 |
| Inspection classified OAI | Feb 3, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1126-2015, this FEI as the recalling firm | Jan 16, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 6, 2013 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |