MEVION MEDICAL SYSTEMS, INC. FDA Inspection Record

FEI 3007087027 · Inspections, Form 483s, warning letters and import refusals

MEVION MEDICAL SYSTEMS, INC. is an FDA-registered facility (FEI 3007087027) located in Littleton, MA, US. RegDataLab tracks 4 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Mar 5, 2020: No Action Indicated (NAI)
Inspections on record
4 (2 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
14, latest Feb 1, 2022

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 5, 2020No Action Indicated (NAI)Failure to submit (21 CFR 1002.13)
Results of tests (21 CFR 1002.30(a)(2))
Mar 30, 2017Voluntary Action Indicated (VAI)Complaints (21 CFR 820.198(a))
Failure to submit (21 CFR 1002.13)
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Mar 16, 2015Voluntary Action Indicated (VAI)Supplier notification of changes (21 CFR 820.50(b))
Failure to submit (21 CFR 1002.13)
Failure to submit, distinct marking (21 CFR 1002.10)
Sep 6, 2012No Action Indicated (NAI)No citations listed

Recalls

Showing the 10 most recent of 14 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Feb 1, 2022Class IIZ-0827-2022Mevion S250 and S250i Proton Therapy Systems hand pendant (P/N 10263 or P/N 37808)
Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is…
Oct 19, 2020Class IIZ-0485-2021MEVION S250i, MEVION S250; Proton Radiation Treatment System - Product Usage: intended to…
Treatment beam information disappears on Treatment Console screen while beam delivery continues potential for harm to a patient could occur
Jun 23, 2020Class IIZ-2686-2020MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation…
The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading…
May 15, 2020Class IIZ-2406-2020MEVION S250i, MEVION S250; Proton Radiation Treatment System
Couch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch…
Oct 1, 2018Class IIZ-0705-2019MEVION S250i
Treatment is allowed to continue (via partials) in situations where dDose1 interlock has tripped
Jun 29, 2018Class IIZ-2746-2018MEVION S250/MEVION S250i Proton Beam Radiation therapy system Product…
QFix has advised Mevion Medical Systems that they are identifying unauthorized combinations of QFix Couch Tops with non-QFix devices in…
Dec 16, 2016Class IIZ-1122-2017MEVION S250-Proton Radiation Therapy Product Usage: Proton Radiation Therapy
An error can occur causing Delta corrections to be lost when one setup field is closed and another is opened
Oct 31, 2016Class IIZ-0409-2017MEVION S250 Product Usage: Proton Radiation Therapy
Completing Setup Without Program Move Complete Can Cause An Isocenter Error
Oct 21, 2016Class IIZ-0410-2017MEVION S250 Product Usage: Proton Radiation Therapy System
Recovery of Incorrect Isocenter -operational change
Oct 20, 2016Class IIZ-0411-2017MEVION S250 Product Usage: Proton Radiation Therapy System
Software defect:2D projection of contours in Verity. The defect causes structures to appear off from their true position in the DRR and the…

FDA record score

10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 26, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

16 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0827-2022, this FEI as the recalling firmFeb 1, 20224 x 0.62.4
Inspection classified VAIMar 30, 20176 x 0.251.5
Class II recall Z-0485-2021, this FEI as the recalling firmOct 19, 20204 x 0.251
Class II recall Z-2686-2020, this FEI as the recalling firmJun 23, 20204 x 0.251
Class II recall Z-2406-2020, this FEI as the recalling firmMay 15, 20204 x 0.251
Class II recall Z-0705-2019, this FEI as the recalling firmOct 1, 20184 x 0.251
Class II recall Z-2746-2018, this FEI as the recalling firmJun 29, 20184 x 0.251
Class II recall Z-1122-2017, this FEI as the recalling firmDec 16, 20164 x 0.251
Class II recall Z-0409-2017, this FEI as the recalling firmOct 31, 201610 years or older, so it adds no points0
Class II recall Z-0410-2017, this FEI as the recalling firmOct 21, 201610 years or older, so it adds no points0
Class II recall Z-0411-2017, this FEI as the recalling firmOct 20, 201610 years or older, so it adds no points0
Class II recall Z-1529-2016, this FEI as the recalling firmMar 15, 201610 years or older, so it adds no points0
Class II recall Z-0767-2016, this FEI as the recalling firmNov 18, 201510 years or older, so it adds no points0
Inspection classified VAIMar 16, 201510 years or older, so it adds no points0
Class II recall Z-0588-2015, this FEI as the recalling firmNov 10, 201410 years or older, so it adds no points0
Class II recall Z-1271-2014, this FEI as the recalling firmMar 5, 201410 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 26, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 26, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 26, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 26, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing MEVION MEDICAL SYSTEMS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check MEVION MEDICAL SYSTEMS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Mevion Medical Systems, Inc.
Location
Littleton, MA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.