MEVION MEDICAL SYSTEMS, INC. FDA Inspection Record
MEVION MEDICAL SYSTEMS, INC. is an FDA-registered facility (FEI 3007087027) located in Littleton, MA, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 5, 2020: No Action Indicated (NAI)
- Inspections on record
- 4 (2 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 14, latest Feb 1, 2022
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 5, 2020 | No Action Indicated (NAI) | Failure to submit (21 CFR 1002.13) Results of tests (21 CFR 1002.30(a)(2)) |
| Mar 30, 2017 | Voluntary Action Indicated (VAI) | Complaints (21 CFR 820.198(a)) Failure to submit (21 CFR 1002.13) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Mar 16, 2015 | Voluntary Action Indicated (VAI) | Supplier notification of changes (21 CFR 820.50(b)) Failure to submit (21 CFR 1002.13) Failure to submit, distinct marking (21 CFR 1002.10) |
| Sep 6, 2012 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 1, 2022 | Class II | Z-0827-2022 | Mevion S250 and S250i Proton Therapy Systems hand pendant (P/N 10263 or P/N 37808) Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is… |
| Oct 19, 2020 | Class II | Z-0485-2021 | MEVION S250i, MEVION S250; Proton Radiation Treatment System - Product Usage: intended to… Treatment beam information disappears on Treatment Console screen while beam delivery continues potential for harm to a patient could occur |
| Jun 23, 2020 | Class II | Z-2686-2020 | MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation… The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading… |
| May 15, 2020 | Class II | Z-2406-2020 | MEVION S250i, MEVION S250; Proton Radiation Treatment System Couch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch… |
| Oct 1, 2018 | Class II | Z-0705-2019 | MEVION S250i Treatment is allowed to continue (via partials) in situations where dDose1 interlock has tripped |
| Jun 29, 2018 | Class II | Z-2746-2018 | MEVION S250/MEVION S250i Proton Beam Radiation therapy system
Product… QFix has advised Mevion Medical Systems that they are identifying unauthorized combinations of QFix Couch Tops with non-QFix devices in… |
| Dec 16, 2016 | Class II | Z-1122-2017 | MEVION S250-Proton Radiation Therapy
Product Usage:
Proton Radiation Therapy An error can occur causing Delta corrections to be lost when one setup field is closed and another is opened |
| Oct 31, 2016 | Class II | Z-0409-2017 | MEVION S250
Product Usage: Proton Radiation Therapy Completing Setup Without Program Move Complete Can Cause An Isocenter Error |
| Oct 21, 2016 | Class II | Z-0410-2017 | MEVION S250
Product Usage: Proton Radiation Therapy System Recovery of Incorrect Isocenter -operational change |
| Oct 20, 2016 | Class II | Z-0411-2017 | MEVION S250
Product Usage: Proton Radiation Therapy System Software defect:2D projection of contours in Verity. The defect causes structures to appear off from their true position in the DRR and the… |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
16 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0827-2022, this FEI as the recalling firm | Feb 1, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Mar 30, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0485-2021, this FEI as the recalling firm | Oct 19, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2686-2020, this FEI as the recalling firm | Jun 23, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2406-2020, this FEI as the recalling firm | May 15, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0705-2019, this FEI as the recalling firm | Oct 1, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2746-2018, this FEI as the recalling firm | Jun 29, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1122-2017, this FEI as the recalling firm | Dec 16, 2016 | 4 x 0.25 | 1 |
| Class II recall Z-0409-2017, this FEI as the recalling firm | Oct 31, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0410-2017, this FEI as the recalling firm | Oct 21, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0411-2017, this FEI as the recalling firm | Oct 20, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1529-2016, this FEI as the recalling firm | Mar 15, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0767-2016, this FEI as the recalling firm | Nov 18, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 16, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0588-2015, this FEI as the recalling firm | Nov 10, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1271-2014, this FEI as the recalling firm | Mar 5, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing MEVION MEDICAL SYSTEMS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check MEVION MEDICAL SYSTEMS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Mevion Medical Systems, Inc.
- Location
- Littleton, MA (US)
- Source
- device