SPECTRANETICS CORP. FDA Inspection Record

FEI 3007284006 · Inspections, Form 483s, warning letters and import refusals

SPECTRANETICS CORP. is an FDA-registered facility (FEI 3007284006) located in Colorado Springs, CO, US. RegDataLab tracks 7 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Mar 1, 2019: No Action Indicated (NAI)
Inspections on record
7 (1 OAI, 2 VAI, 4 NAI)
Last OAI inspection
Jan 21, 2016
Warning letters
1 listed by FDA for this FEI, latest May 23, 2016
Recalls as the recalling firm
10, latest Jan 10, 2025

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 1, 2019No Action Indicated (NAI)No citations listed
Sep 27, 2017Voluntary Action Indicated (VAI)Training - Lack of or inadequate procedures (21 CFR 820.25(b))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)
Environmental control Lack of or inadequate procedures (21 CFR 820.70(c))
Lack of or inadequate process validation (21 CFR 820.75(a))
Jan 21, 2016Official Action Indicated (OAI)No citations listed
Mar 14, 2013No Action Indicated (NAI)No citations listed
Nov 16, 2011No Action Indicated (NAI)No citations listed
Nov 5, 2010Voluntary Action Indicated (VAI)No citations listed
Feb 26, 2009No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jan 10, 2025Class IZ-1070-2025Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack…
Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner…
Nov 25, 2024Class IIZ-0680-2025Turbo Elite Laser Atherectomy Catheter Model Number / UDI-DI code / Product…
Due to complaints received for the marker band detaching from the distal tip of the catheter and remaining in the patient as laser catheter…
Nov 15, 2023Class IIZ-0509-2024Spectranetics Turbo-Elite Laser Atherectomy Catheter 2.0 RX REF 420-159 Spectranetics…
Incorrect product labeling. Exterior product box label does not match internal pouch label.
Oct 14, 2022Class IIZ-0221-2023Philips Laser System, REF: LAS-100. Used in minimally invasive interventional procedures…
Laser System may detect Error 106 or Error 108 System Failures that will cause the system to enter non-recoverable safe state. The user will…
Jun 1, 2022Class IIZ-1351-2022Philips Laser System, REF LAS-100, Intermittent Operation, Duty cycle dependent upon…
The LAS-100 Laser system may detect an inoperable hardware component during power up, which results in an error code and the system not…
Mar 31, 2020Class IIZ-2143-2020Bridge Occlusion Balloon, REF 590-001, QTY 1, Sheath Compatibility 12F, Guidewire…
There is a potential for thrombus formation on balloons prepped and staged in the vasculature for long durations. All users, such as those…
Aug 7, 2017Class IZ-3106-2017Spectranetics Bridge Occlusion Balloon catheter, Sheath Compatibility 12F, Guidewire…
Possible inability to pass the guidewire through the Bridge device, therefore, resulting in a potential delay of treatment.
Jun 6, 2017Class IIZ-2863-2017Spectranetics Turbo-Elite Laser Atherectomy Catheter Peripheral Rapid Exchange Catheters…
Potential for flouroscopially visible marker band at the tip of the Rapid Exchange Turbo Elite Atherectomy Catheters to become loose or…
Oct 31, 2016Class IIZ-0661-2017Turbo Elite Atherectomy Catheter. Models: 410-152, 410-154, 414-151, 414-159, 417-152… and 2 more products
Spectranetics Corporation announces a voluntary field action for the Turbo Elite Atherectomy Catheter, ELCA Coronary Atherectomy Catheter…
Jan 15, 2016Class IIZ-0758-20162.3mm Turbo-Power OTW, Manual, Model No 423-050 Product Usage The Turbo-Power System…
The product is labeled with an expiration date that is past its shelf life.

FDA record score

34 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

14 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-1070-2025, this FEI as the recalling firmJan 10, 202515 x 115
Class II recall Z-0680-2025, this FEI as the recalling firmNov 25, 20244 x 14
Class I recall Z-3106-2017, this FEI as the recalling firmAug 7, 201715 x 0.253.8
Class II recall Z-0509-2024, this FEI as the recalling firmNov 15, 20234 x 0.62.4
Class II recall Z-0221-2023, this FEI as the recalling firmOct 14, 20224 x 0.62.4
Class II recall Z-1351-2022, this FEI as the recalling firmJun 1, 20224 x 0.62.4
Inspection classified VAISep 27, 20176 x 0.251.5
Class II recall Z-2143-2020, this FEI as the recalling firmMar 31, 20204 x 0.251
Class II recall Z-2863-2017, this FEI as the recalling firmJun 6, 20174 x 0.251
Class II recall Z-0661-2017, this FEI as the recalling firmOct 31, 201610 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsMay 23, 201610 years or older, so it adds no points0
Inspection classified OAIJan 21, 201610 years or older, so it adds no points0
Class II recall Z-0758-2016, this FEI as the recalling firmJan 15, 201610 years or older, so it adds no points0
Inspection classified VAINov 5, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing SPECTRANETICS CORP. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check SPECTRANETICS CORP. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Spectranetics Corporation
Location
Colorado Springs, CO (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.