Extremity Medical, LLC FDA Inspection Record

FEI 3007289093 · Inspections, Form 483s, warning letters and import refusals

Extremity Medical, LLC is an FDA-registered facility (FEI 3007289093) located in Parsippany, NJ, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Jun 12, 2026: No Action Indicated (NAI)
Inspections on record
6 (1 OAI, 1 VAI, 4 NAI)
Last OAI inspection
Oct 24, 2011
Warning letters
1 listed by FDA for this FEI, latest Feb 24, 2012
Recalls as the recalling firm
4, latest Jan 10, 2013

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jun 12, 2026No Action Indicated (NAI)No citations listed
May 20, 2024No Action Indicated (NAI)No citations listed
Oct 17, 2019No Action Indicated (NAI)No citations listed
May 24, 2016No Action Indicated (NAI)No citations listed
Oct 9, 2012Voluntary Action Indicated (VAI)Lack of or inadequate DHR procedures (21 CFR 820.184)
Report of risk to health (21 CFR 806.10(a)(1))
Oct 24, 2011Official Action Indicated (OAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Nonconforming product control (21 CFR 820.90(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Investigation of device failures (21 CFR 820.198(c))
and 2 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jan 10, 2013Class IIZ-0952-2013Extremity Medical - Disposable Orthopedic Surgical Instrument - 1.6 mm Guidewire…
The 1.6 mm Guidewire is undersized up to 2 mm in length. When used with the depth gauge, the measurement reading will not correspond with…
Mar 22, 2012Class IIZ-0752-2013Extremity Medical, IO FiX System Accessory. Catalog Number 101-00010. This screw…
Lag screw measurement is short (approximately 5 mm) when using current screw measuring device (p/n 101-00010) per the IO FIX Surgical…
Apr 25, 2011Class IIZ-0438-2013IO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service…
Product Event Reports revealed a broken reamer experienced intra-operatively. Pre-drilling was not performed prior to reaming.
Apr 13, 2011Class IIZ-0344-2013IO Fix IntraOsseous Fixation System; Extremity Medical. www.extremitymedical.com…
Product Event Reports revealed that lag screw passes through washer (e.g. X-Post) intra-operatively.

FDA record score

0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 16, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0952-2013, this FEI as the recalling firmJan 10, 201310 years or older, so it adds no points0
Inspection classified VAIOct 9, 201210 years or older, so it adds no points0
Class II recall Z-0752-2013, this FEI as the recalling firmMar 22, 201210 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsFeb 24, 201210 years or older, so it adds no points0
Inspection classified OAIOct 24, 201110 years or older, so it adds no points0
Class II recall Z-0438-2013, this FEI as the recalling firmApr 25, 201110 years or older, so it adds no points0
Class II recall Z-0344-2013, this FEI as the recalling firmApr 13, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 16, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 16, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 16, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 16, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing Extremity Medical, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.