Trilliant Surgical LLC FDA Inspection Record
Trilliant Surgical LLC is an FDA-registered facility (FEI 3007420745) located in Houston, TX, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Dec 11, 2019: Voluntary Action Indicated (VAI)
- Inspections on record
- 6 (3 OAI, 3 VAI)
- Last OAI inspection
- Jun 27, 2016
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest May 3, 2010. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 13, latest Nov 4, 2019
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Dec 11, 2019 | Voluntary Action Indicated (VAI) | Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h)) Documentation (21 CFR 820.75(a)) Investigation of device failures (21 CFR 820.198(c)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) and 1 more |
| May 2, 2018 | Voluntary Action Indicated (VAI) | Design output - review and approval (21 CFR 820.30(d)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate In-process acceptance procedures (21 CFR 820.80(c)) Not approved or obsolete document retrieval (21 CFR 820.40(a)) and 6 more |
| Jun 27, 2016 | Official Action Indicated (OAI) | Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Documentation (21 CFR 820.75(a)) Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a)) and 5 more |
| Feb 12, 2015 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 1 more |
| Sep 9, 2011 | Voluntary Action Indicated (VAI) | Quality audits - auditor independence (21 CFR 820.22) Quality Audit/Reaudit - conducted (21 CFR 820.22) Documentation (21 CFR 820.100(b)) |
| Jan 20, 2010 | Official Action Indicated (OAI) | Approval, inadequate purchasing data (21 CFR 820.50(b)) CAPA verification/validation of corrective/preventive action (21 CFR 820.100(a)(4)) Documentation (21 CFR 820.100(b)) Report of risk to health (21 CFR 806.10(a)(1)) |
Warning letters
- May 3, 2010: Warning letter listed on FDA's Data Dashboard. View compliance record (108307).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Nov 4, 2019 | Class II | Z-1635-2020 | 8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only… The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile… |
| Oct 10, 2019 | Class II | Z-1892-2020 | 3.5mm Aresenal Screw Drill Bit , Ref: 330-35-001, Qty. 70, Rx Only, Non-Sterile, UDI… Labeling Error; specifically, drill bits from lots # TSL008495A are marked with the lot # of TSL008495. This could lead to parts being… |
| Mar 14, 2019 | Class II | Z-0799-2020 | Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates… A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive… |
| Jul 19, 2018 | Class II | Z-1605-2020 | Instructions for Use and Surgical Cleaning and Sterilization Protocol for:
(1)… and 9 more products Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU… |
| Mar 27, 2018 | Class II | Z-2338-2018 | Cannulated Driver Handle, Catalog #: 210-00-003, Rx Only, Non Sterile and 1 more product Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the… |
| Dec 8, 2016 | Class II | Z-0960-2017 | Tiger Screw Sterilization Tray Base Assembly Model 210-00-001 Revision N, O
The product… Four lots of Tiger Screw Sterilization Tray Base Assemblies were released to the field with a printing error where k-wire description… |
| Sep 2, 2016 | Class II | Z-0609-2017 | Model 208-70-110, Long Thread Tiger Large Cannulated Headless Screw, 7 x 110 mm labeled… Screws were identified as having the incorrect part number and lot number marked upon them. The part number incorrectly indicated a length… |
| Aug 6, 2015 | Class II | Z-2490-2015 | 7 Hole VL Gridlock Fibula Plate, Part # 300-60-001 NON-STERILE SINGLE USE ONLY; 9 Hole VL… The affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of… |
| Jul 31, 2015 | Class II | Z-2738-2015 | 2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number… Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow… |
| Jul 20, 2015 | Class II | Z-2421-2015 | 2.0/2.4 Cannulated Screw Countersink Part # 210-24-002 NON-STERILE and 3.0/4.0 Cannulated… The cutting efficiency of the Countersinks was out of specification. |
FDA record score
9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
20 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Dec 11, 2019 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | May 2, 2018 | 6 x 0.25 | 1.5 |
| Class II recall Z-1635-2020, this FEI as the recalling firm | Nov 4, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1892-2020, this FEI as the recalling firm | Oct 10, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0799-2020, this FEI as the recalling firm | Mar 14, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1605-2020, this FEI as the recalling firm | Jul 19, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2338-2018, this FEI as the recalling firm | Mar 27, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0960-2017, this FEI as the recalling firm | Dec 8, 2016 | 4 x 0.25 | 1 |
| Class II recall Z-0609-2017, this FEI as the recalling firm | Sep 2, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Jun 27, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2490-2015, this FEI as the recalling firm | Aug 6, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2738-2015, this FEI as the recalling firm | Jul 31, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2421-2015, this FEI as the recalling firm | Jul 20, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2452-2015, this FEI as the recalling firm | Jul 20, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2456-2015, this FEI as the recalling firm | Jul 20, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2464-2015, this FEI as the recalling firm | Jul 20, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Feb 12, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Sep 9, 2011 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | May 3, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Jan 20, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 27, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 27, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 27, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 27, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Trilliant Surgical LLC as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Trilliant Surgical LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Trilliant Surgical LLC → Check this supplier
Facility details
- Company
- Trilliant Surgical, LLC
- Location
- Houston, TX (US)
- Source
- device