THERMEDX, LLC FDA Inspection Record
THERMEDX, LLC is an FDA-registered facility (FEI 3007495879) located in Solon, OH, US. RegDataLab tracks 2 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jul 28, 2016: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (1 OAI, 1 VAI)
- Last OAI inspection
- Dec 11, 2014
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Apr 8, 2015. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Sep 1, 2014
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 28, 2016 | Voluntary Action Indicated (VAI) | Nonconforming product control (21 CFR 820.90(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Dec 11, 2014 | Official Action Indicated (OAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Lack of or inadequate procedures (21 CFR 820.100(a)) DHR content (21 CFR 820.184) Supplier oversight (21 CFR 820.50(a)(2)) and 2 more |
Warning letters
- Apr 8, 2015: Warning letter listed on FDA's Data Dashboard. View compliance record (448762).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 1, 2014 | Class II | Z-1463-2015 | Fluid Management System P4000; for irrigation and fluid warming in laparoscopic… To correct software bugs that could affect the ability to accurately measure fluid deficit. |
| Feb 10, 2014 | Class II | Z-2555-2014 | The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate… A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring. The… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jul 28, 2016 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Apr 8, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Dec 11, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1463-2015, this FEI as the recalling firm | Sep 1, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2555-2014, this FEI as the recalling firm | Feb 10, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing THERMEDX, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check THERMEDX, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Thermedx LLC
- Location
- Solon, OH (US)
- Source
- device