Drive DeVilbiss Healthcare FDA Inspection Record
Drive DeVilbiss Healthcare is an FDA-inspected facility (FEI 3007538326) located in Canfield, OH, US. RegDataLab tracks 3 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Feb 24, 2016: No Action Indicated (NAI)
- Inspections on record
- 3 (1 OAI, 2 NAI)
- Last OAI inspection
- Aug 31, 2012
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Oct 10, 2012. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Sep 15, 2014
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 24, 2016 | No Action Indicated (NAI) | No citations listed |
| Jul 18, 2014 | No Action Indicated (NAI) | Nonconforming product control (21 CFR 820.90(a)) Investigation of device failures (21 CFR 820.198(c)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) |
| Aug 31, 2012 | Official Action Indicated (OAI) | Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3)) Lack of or inadequate procedures (21 CFR 820.100(a)) Complaints (21 CFR 820.198(a)) Design validation - production units (21 CFR 820.30(g)) and 4 more |
Warning letters
- Oct 10, 2012: Warning letter listed on FDA's Data Dashboard. View compliance record (364118).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 15, 2014 | Class II | Z-0067-2015 | AC Powered adjustable bed frame.
AC powered adjustable bed frame used in nursing… Firm received several complaints that the head deck actuator bracket had separated from the bed frame. |
| Aug 11, 2014 | Class II | Z-2596-2014 | Prime Care ¿ Transcend Mattress, Prime Care ¿ Transcend Perimeter Mattress, Prime Care ¿… The firm found that the top cover of the mattresses were delaminating. |
| Feb 2, 2011 | Class II | Z-0270-2013 | AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC… The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0067-2015, this FEI as the recalling firm | Sep 15, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2596-2014, this FEI as the recalling firm | Aug 11, 2014 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Oct 10, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Aug 31, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0270-2013, this FEI as the recalling firm | Feb 2, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 27, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 27, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 27, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 27, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Drive DeVilbiss Healthcare as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Drive DeVilbiss Healthcare: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Drive DeVilbiss Healthcare → Check this supplier
Facility details
- Company
- Drive DeVilbiss Healthcare
- Location
- Canfield, OH (US)
- Source
- inspections