Dutch Ophthalmic USA, Inc. FDA Inspection Record
Dutch Ophthalmic USA, Inc. is an FDA-inspected facility (FEI 3007916166) located in Exeter, NH, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Oct 24, 2016: No Action Indicated (NAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Apr 14, 2021
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 24, 2016 | No Action Indicated (NAI) | No citations listed |
| Jun 3, 2011 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Evaluation, timeliness, identification (21 CFR 820.198(d)) Lack of or inadequate organizational structure (21 CFR 820.20(b)) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Apr 14, 2021 | Class II | Z-2097-2021 | Illuminated Stepped Laser Probe Curved-for use in combination with 23G vitrectomy… Barcode label extended expiration date. Barcode located at the short side of the outer box shows 2025-10-27 when scanned, while the correct… |
| Feb 25, 2021 | Class II | Z-2091-2021 | Disposable Replacement Backflush Brush Needles (23 gauge/0.6mm)-for use in vitreoretinal… Mislabeled: 1281.B1D Disposable Replacement Backflush Brush Needles label states (20 gauge/0.9 mm) instead of the correct size of 23… |
| May 21, 2019 | Class II | Z-2252-2021 | The device is a 25-gauge infusion line that is part of 25-gauge surgical pack labeled TDC… Labeled 25-gauge surgical pack, TDC Vitrectomy Pack VGPC 25g, containing 23-gauge infusion line |
FDA record score
3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2097-2021, this FEI as the recalling firm | Apr 14, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2091-2021, this FEI as the recalling firm | Feb 25, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2252-2021, this FEI as the recalling firm | May 21, 2019 | 4 x 0.25 | 1 |
| Inspection classified VAI | Jun 3, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 29, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 29, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 29, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 29, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Dutch Ophthalmic USA, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Dutch Ophthalmic USA, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Dutch Ophthalmic USA, Inc. → Check this supplier
Facility details
- Company
- Dutch Ophthalmic USA, Inc.
- Location
- Exeter, NH (US)
- Source
- inspections