Spinal Elements FDA Inspection Record

FEI 3007970248 · Inspections, Form 483s, warning letters and import refusals

Spinal Elements is an FDA-inspected facility (FEI 3007970248) located in Marietta, GA, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Sep 28, 2017: Voluntary Action Indicated (VAI)
Inspections on record
5 (1 VAI, 4 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
10, latest Apr 16, 2018

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 28, 2017Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Lack of procedures, or not maintained (21 CFR 820.40)
Feb 14, 2017No Action Indicated (NAI)No citations listed
Nov 12, 2014No Action Indicated (NAI)No citations listed
May 14, 2013No Action Indicated (NAI)No citations listed
Jan 5, 2011No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Apr 16, 2018Class IIZ-1912-2018AMENDIA Omega Lumbar Interbody Fusion Device, Part Number 72-00-2-092812-11 The…
Omega LIF interbody implants labeled as having 11 degrees of lordosis was assembled using components manufactured with 4 degrees of…
Nov 13, 2017Class IIZ-0239-2018Piranha Screw 4.0 x 12mm, used in Piranha Surgical Plate System The Piranha Surgical…
Potential incorrect marking and color identification of Piranha screw. Some screws marked as 12 mm actually measure 14 mm.
Jul 7, 2017Class IIZ-0316-2018Ceres Self-Drilling Variable Screw 4.0 x 12mm The Ceres screws are self-drilling…
Incorrect direct marking and color identification for the affected lot of Ceres Self-Drilling Variable Screws. Although the laser mark and…
Jul 7, 2017Class IIZ-0325-2018Mac Pin Non-Cannulated Screw, 6.5 x 60mm
Incorrect labeling for the lot of MAC Pins. Although the pins are laser as 460 mm in length, the pins are 60 mm in length.
Apr 6, 2017Class IIZ-1885-201701.4mm x 457mm (18in) K-Wire, Threaded Blunt, Ref 9080B-18T, Qty: 1 per pack, Material…
Mislabeled guide wires - It was discovered that some Guide Wires labeled as stainless steel were actually manufactured from nitinol.
Nov 22, 2016Class IIZ-0827-2017Zeus-P Lumbar Interbody Device
Labeling error where three boxes in inventory had conflicting patient and primary labels. Evaluation indicated that the primary label and…
Nov 22, 2016Class IIZ-0881-2017K-Wire 1.4mm x 12in Non-Threaded, Part Number 9080-12U; K-Wire Threaded 1.4mm x 18, Part… and 2 more products
The rods are packaged in packages with unacceptable seals. In some cases, the integrity of the seal is compromised by channels or voids.
Nov 22, 2016Class IIZ-0886-2017Black Widow Plate 19mm Plate, Part Number TBP119; Black Widow Plate 23mm Plate, Part… and 1 more product
There is a problem with the mating feature between the plate and screw which could cause the plate opening not to interfere enough with the…
Nov 22, 2016Class IIZ-0891-2017Optimus Fixed Awl
Complaints that the tip of the awl broke after impaction during surgery.
Nov 22, 2016Class IIZ-0966-2017Ceres-C Standalone Cervical Interbody The product is a stand-alone anterior cervical…
Absent tantalum market pin.

FDA record score

12 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

11 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAISep 28, 20176 x 0.251.5
Class II recall Z-1912-2018, this FEI as the recalling firmApr 16, 20184 x 0.251
Class II recall Z-0239-2018, this FEI as the recalling firmNov 13, 20174 x 0.251
Class II recall Z-0316-2018, this FEI as the recalling firmJul 7, 20174 x 0.251
Class II recall Z-0325-2018, this FEI as the recalling firmJul 7, 20174 x 0.251
Class II recall Z-1885-2017, this FEI as the recalling firmApr 6, 20174 x 0.251
Class II recall Z-0827-2017, this FEI as the recalling firmNov 22, 20164 x 0.251
Class II recall Z-0881-2017, this FEI as the recalling firmNov 22, 20164 x 0.251
Class II recall Z-0886-2017, this FEI as the recalling firmNov 22, 20164 x 0.251
Class II recall Z-0891-2017, this FEI as the recalling firmNov 22, 20164 x 0.251
Class II recall Z-0966-2017, this FEI as the recalling firmNov 22, 20164 x 0.251

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Spinal Elements as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Spinal Elements yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Spinal Elements
Location
Marietta, GA (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.