Latest inspection
Apr 21, 2015: Official Action Indicated (OAI)
Inspections on record
2 (1 OAI, 1 NAI)
Last OAI inspection
Apr 21, 2015
Warning letters
2 listed by FDA for this FEI, latest Jan 1, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Apr 21, 2015Official Action Indicated (OAI)Design verification - output does not meet input requirement (21 CFR 820.30(f))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Lack of or inadequate procedures (21 CFR 820.100(a))
Design validation - user needs and intended uses (21 CFR 820.30(g))
and 2 more
Jul 15, 2014No Action Indicated (NAI)No citations listed