OMNIlifeScience, Inc. FDA Inspection Record
OMNIlifeScience, Inc. is an FDA-registered facility (FEI 3008342029) located in Raynham, MA, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Apr 4, 2022: No Action Indicated (NAI)
- Inspections on record
- 4 (2 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 13, latest Oct 1, 2021
- Import refusals
- 3
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 4, 2022 | No Action Indicated (NAI) | No citations listed |
| Sep 13, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Management review - Lack of or inadequate procedures (21 CFR 820.20(c)) |
| Dec 6, 2012 | Voluntary Action Indicated (VAI) | Investigation of device failures (21 CFR 820.198(c)) Management review - Lack of or inadequate procedures (21 CFR 820.20(c)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Feb 3, 2011 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 1, 2021 | Class II | Z-0237-2022 | OMNI K2 Hip Stem, Manufacturer's Product Code 130006
The OMNI K2 Hip Stem is intended for… The device is missing grit blasting on the proximal surface of the femoral stem, which may lead to the possibility of product revision in… |
| Dec 21, 2020 | Class II | Z-0920-2021 | SERF Novae Dual Mobility Cup Insert 51/28- ApeX-LNK Poly Liner. Intended for use with the… and 2 more products Acetabular Liners have a higher than acceptable oxidation limit, potential for patients to experience a higher wear rate for acetabular… |
| Sep 11, 2019 | Class II | Z-0169-2020 | OMNI Hip System
Model, Catalogue, or Order Number(s): HC-00070
Intended for use as the… Stems missing porous coating on the device. |
| Jul 23, 2019 | Class II | Z-2479-2019 | OMNIlife science, Inc.
ApeX acetabular liner
hip prosthesis component
Serf Dual… and 2 more products Acetabular liners failed the oxidation testing performed prior to release. They may experience a higher wear rate, leading to revision… |
| Apr 12, 2019 | Class II | Z-1717-2019 | OMNIBotics iBlock Cutting Guide, 4146-1000 - Product Usage: The OMNIBotics iBlock Cutting… The OMNIBotics iBlock Cutting Guide does not lock onto the iBlock Saw Guide as intended and may loosen during use. |
| Nov 29, 2018 | Class II | Z-0837-2019 | OMNI K1 Broach Handle, Product Code HS-10054 There is a potential for the inner pin of the handle to become loose and fall out. |
| May 18, 2018 | Class II | Z-2540-2018 | OMNI Hip System, ApeX-LNK Acetabular Inserts 0 degree, 4+mm offset, Product Codes… The Internal packaging, the pouch and patient labels are improperly labeled. |
| Nov 20, 2017 | Class II | Z-1117-2018 | Apex Knee System, Tibial Baseplate Size 6 Left, non-porous - KC-2206L
The Apex Knee… The Apex Knee System Tibial Baseplate identified has an improper expiration date marked on the external packaging. The improper expiration… |
| Jun 9, 2015 | Class II | Z-2454-2015 | Offset Femoral Alignment Guide, 2mm x 6¿; Product Code: KS 67022
Product Usage:
The… and 1 more product The markings on both versions of the OFFSET FEMORAL ALIGNMENT GUIDES are incorrect. Specifically, the markings on the alignment guides are… |
| Jun 4, 2015 | Class II | Z-2445-2015 | APEX INTERFACE Hemi 54X and Flared 56X, 3 Hole Acetabular Shells; Model: H5-12354… The device may have improper screw hole placement due to inaccurate location of the index line. |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
15 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Sep 13, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0237-2022, this FEI as the recalling firm | Oct 1, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0920-2021, this FEI as the recalling firm | Dec 21, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0169-2020, this FEI as the recalling firm | Sep 11, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2479-2019, this FEI as the recalling firm | Jul 23, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1717-2019, this FEI as the recalling firm | Apr 12, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0837-2019, this FEI as the recalling firm | Nov 29, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2540-2018, this FEI as the recalling firm | May 18, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1117-2018, this FEI as the recalling firm | Nov 20, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2454-2015, this FEI as the recalling firm | Jun 9, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2445-2015, this FEI as the recalling firm | Jun 4, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2064-2015, this FEI as the recalling firm | Jun 3, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2241-2014, this FEI as the recalling firm | May 9, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Dec 6, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2370-2012, this FEI as the recalling firm | Aug 9, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing OMNIlifeScience, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check OMNIlifeScience, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- OmniLife Science
- Location
- Raynham, MA (US)
- Source
- device