MAX MOBILITY, LLC FDA Inspection Record

FEI 3008370857 · Inspections, Form 483s, warning letters and import refusals

MAX MOBILITY, LLC is an FDA-registered facility (FEI 3008370857) located in Lebanon, TN, US. RegDataLab tracks 4 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Dec 7, 2023: Voluntary Action Indicated (VAI)
Inspections on record
4 (1 OAI, 3 VAI)
Last OAI inspection
Mar 27, 2014
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
3, latest Aug 12, 2025

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Dec 7, 2023Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Nov 7, 2017Voluntary Action Indicated (VAI)Design verification - documentation (21 CFR 820.30(f))
Apr 24, 2015Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of Written MDR Procedures (21 CFR 803.17)
Mar 27, 2014Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design validation - software validation documentation (21 CFR 820.30(g))
Design review - Lack of or inadequate procedures (21 CFR 820.30(e))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
and 7 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Aug 12, 2025Class IZ-2538-2025Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DC use in conjunction… and 2 more products
Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of…
Dec 20, 2024Class IZ-1116-2025Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DC use in conjunction… and 1 more product
Due to faulty circuit board in speed control remote use in conjunction with a wheelchair power assist, their is a potential for lost of…
May 10, 2023Class IIZ-2004-2023PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part… and 1 more product
When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens…

FDA record score

38 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-2538-2025, this FEI as the recalling firmAug 12, 202515 x 115
Class I recall Z-1116-2025, this FEI as the recalling firmDec 20, 202415 x 115
Inspection classified VAIDec 7, 20236 x 0.63.6
Class II recall Z-2004-2023, this FEI as the recalling firmMay 10, 20234 x 0.62.4
Inspection classified VAINov 7, 20176 x 0.251.5
Inspection classified VAIApr 24, 201510 years or older, so it adds no points0
Inspection classified OAIMar 27, 201410 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing MAX MOBILITY, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check MAX MOBILITY, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Max Mobility LLC
Location
Lebanon, TN (US)
Source