Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 10, 2023 | No Action Indicated (NAI) | No citations listed |
| Jan 16, 2020 | No Action Indicated (NAI) | No citations listed |
| Sep 8, 2017 | Voluntary Action Indicated (VAI) | Report of risk to health (21 CFR 806.10(a)(1)) |
| Jun 24, 2014 | No Action Indicated (NAI) | No citations listed |
| Jun 5, 2013 | No Action Indicated (NAI) | No citations listed |
| Apr 2, 2012 | Official Action Indicated (OAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Design validation - user needs and intended uses (21 CFR 820.30(g)) Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2)) Documentation (21 CFR 820.100(b)) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Nov 5, 2015 | Class II | Z-0040-2018 | Genesys spine, Binary Screwdriver, GCP221, component of the Binary Anterior Cervical… Unanticipated technique for a cervical screw insertion caused the tip of the threaded inner shaft to break off. |
| Apr 2, 2014 | Class II | Z-2677-2014 | Circular Lock Screws, part # G826, lot # 11406. The Lock Screws are part of the Genesys… The recall is being initiated because MK Precision (the Circular Lock Screw supplier) notified Genesys Spine that a portion (less than 20%)… |
| Feb 20, 2013 | Class II | Z-0921-2013 | 3.75mm x 12mm Fixed Cervical Screws. Components of the Genesys Spine Anterior Cervical… Manufacturing residue may be present on the cervical screws. Use of this product may result in patient infection, delayed fusion or… |
FDA record score
2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Sep 8, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0040-2018, this FEI as the recalling firm | Nov 5, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2677-2014, this FEI as the recalling firm | Apr 2, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0921-2013, this FEI as the recalling firm | Feb 20, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Apr 2, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing GENESYS SPINE as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for GENESYS SPINE: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Facility details
- Company
- Genesys Orthopedics Systems, LLC
- Location
- Austin, TX (US)
- Source
- device