SurgiSmoke Solutions FDA Inspection Record
SurgiSmoke Solutions is an FDA-registered facility (FEI 3008493151) located in Saint Paul, MN, US. RegDataLab tracks 6 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Feb 23, 2026: No Action Indicated (NAI)
- Inspections on record
- 6 (6 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Mar 16, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 23, 2026 | No Action Indicated (NAI) | Design and development validation not complete prior to release of product (ISO 13485:2016 Clause 7.3.7) Product release and delivery proceeded prior to planned agreements completed (ISO 13485:2016 Clause 8.2.6) Corrective action procedures not documented (ISO 13485:2016 Clause 8.5.2) Integrity of the QMS not maintained (ISO 13485:2016 Clause 5.4.2(b)) |
| Nov 17, 2022 | No Action Indicated (NAI) | No citations listed |
| Aug 8, 2019 | No Action Indicated (NAI) | No citations listed |
| Feb 27, 2015 | No Action Indicated (NAI) | No citations listed |
| Jul 27, 2011 | No Action Indicated (NAI) | No citations listed |
| Feb 13, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Mar 16, 2026 | Class II | Z-1924-2026 | Flamingo Funnel Small, Model Number SQ20012-01 and 2 more products Certain lots of product were associated with a manufacturing nonconformance involving use of a non-gamma sterilization compatible drape. |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
1 record behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1924-2026, this FEI as the recalling firm | Mar 16, 2026 | 4 x 1 | 4 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 17, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 17, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 17, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 17, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 17, 2026 | Newest record Nov 16, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing SurgiSmoke Solutions as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.