SENSUS HEALTHCARE FDA Inspection Record
SENSUS HEALTHCARE is an FDA-registered facility (FEI 3008513398) located in Boca Raton, FL, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jun 27, 2024: No Action Indicated (NAI)
- Inspections on record
- 4 (1 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Feb 10, 2023
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jun 27, 2024 | No Action Indicated (NAI) | No citations listed |
| Aug 5, 2022 | No Action Indicated (NAI) | Design verification - documentation (21 CFR 820.30(f)) |
| Jul 11, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Records of complaint investigation (21 CFR 820.198(e)) |
| Jul 29, 2011 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 10, 2023 | Class II | Z-1235-2023 | Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version… When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient. |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
2 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1235-2023, this FEI as the recalling firm | Feb 10, 2023 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Jul 11, 2017 | 6 x 0.25 | 1.5 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing SENSUS HEALTHCARE as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check SENSUS HEALTHCARE yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Sensus Healthcare, Inc.
- Location
- Boca Raton, FL (US)
- Source