Daheng New Epoch Technology, Inc. FDA Inspection Record
Daheng New Epoch Technology, Inc. is an FDA-inspected facility (FEI 3008705657) located in Beijing, CN. RegDataLab tracks 3 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jul 14, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jan 15, 2014. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Jul 15, 2013
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 14, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) |
| May 20, 2014 | No Action Indicated (NAI) | No citations listed |
| Jul 19, 2013 | Voluntary Action Indicated (VAI) | Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) ID label, labeling (21 CFR 820.184(e)) Lack of System for Event Evaluations (21 CFR 803.17(a)(1)) |
Warning letters
- Jan 15, 2014: Warning letter listed on FDA's Data Dashboard. View compliance record (411469).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 15, 2013 | Class II | Z-1145-2014 | Denlase: Diode Laser Therapy System and Penlase:
Dental Laser Therapy System surgical… Failure to provide calibration procedures to purchasers, as required by 21 CFR 1040.11(a)(2). For example, the user manual did not contain… |
FDA record score
2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jul 14, 2017 | 6 x 0.25 | 1.5 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jan 15, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 19, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1145-2014, this FEI as the recalling firm | Jul 15, 2013 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 28, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 28, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 28, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 27, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Daheng New Epoch Technology, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Daheng New Epoch Technology, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Daheng New Epoch Technology, Inc. → Check this supplier
Facility details
- Company
- Daheng New Epoch Technology, Inc.
- Location
- Beijing (CN)
- Source
- inspections