SUMMIT MEDICAL LLC FDA Inspection Record

FEI 3008719017 · Inspections, Form 483s, warning letters and import refusals

SUMMIT MEDICAL LLC is an FDA-registered facility (FEI 3008719017) located in Saint Paul, MN, US. RegDataLab tracks 8 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
May 14, 2018: Voluntary Action Indicated (VAI)
Inspections on record
8 (2 VAI, 6 NAI)
Last OAI inspection
None on record
Warning letters
1 listed by FDA for this FEI, latest Apr 25, 2013
Recalls as the recalling firm
4, latest Jan 13, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
May 14, 2018Voluntary Action Indicated (VAI)Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
Equipment suitability & capability (21 CFR 820.72(a))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)
and 2 more
Mar 9, 2018Voluntary Action Indicated (VAI)Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
Lack of or inadequate procedures (21 CFR 820.100(a))
Lack of or inadequate process validation (21 CFR 820.75(a))
Nov 10, 2016No Action Indicated (NAI)Design output - Lack of or inadequate procedures (21 CFR 820.30(d))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Documentation (21 CFR 820.80(e))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 1 more
Mar 4, 2015No Action Indicated (NAI)No citations listed
May 5, 2014No Action Indicated (NAI)Documentation (21 CFR 820.75(a))
Nonconforming product control (21 CFR 820.90(a))
Documentation (21 CFR 820.100(b))
Jan 28, 2013No Action Indicated (NAI)Design verification - output does not meet input requirement (21 CFR 820.30(f))
Design review - Lack of or inadequate procedures (21 CFR 820.30(e))
Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 7 more
Sep 7, 2012No Action Indicated (NAI)Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
Lack of or inadequate In-process acceptance procedures (21 CFR 820.80(c))
Nov 6, 2009No Action Indicated (NAI)Quantity manufactured (21 CFR 820.184(b))
Equipment control activity documentation (21 CFR 820.72(a))
Documentation (21 CFR 820.80(c))
Nonconforming product evaluation/investigation (21 CFR 820.90(a))
and 1 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jan 13, 2023Class IIZ-1171-2023Summit Medical Nasal Septal Button, Medium, 5cm, REF SP-78105, sterile.
Sterile product pouches were not sealed.
Sep 3, 2015Class IIZ-0435-2016AbsorbENT, Sterile Compressed Absorbent Sponge, Nasal Pack, 8.9cm(L) x 3.0cm (W) x 1.5cm…
Summit Medical is recalling Doyle P/F Nasal Pack w/Airway, string 8 cm x 3 cm x 1.5 cm, Catalog # RH-7413-10, Lot # 160889 & 161354 because…
May 12, 2014Class IIZ-1731-2014Armstrong Grommet with wire Otological Ventilation Tube Catalog #: VT-0507-01…
One lot of ventilation tubes have mixed tubes with wire attached to the slanted flange and ventilation tubes with wire attached to the…
Mar 27, 2013Class IIZ-1319-2013Summit Medical, BL-2006, Myringotomy Lance (Juvenile) Blade, Sterile P, Summit Medical…
Summit Medical has initiated a recall of Myringotomy Lance Blade (Juvenile) because the box contains the wrong blade. The box contains…

FDA record score

5 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 15, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1171-2023, this FEI as the recalling firmJan 13, 20234 x 0.62.4
Inspection classified VAIMay 14, 20186 x 0.251.5
Inspection classified VAIMar 9, 20186 x 0.251.5
Class II recall Z-0435-2016, this FEI as the recalling firmSep 3, 201510 years or older, so it adds no points0
Class II recall Z-1731-2014, this FEI as the recalling firmMay 12, 201410 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsApr 25, 201310 years or older, so it adds no points0
Class II recall Z-1319-2013, this FEI as the recalling firmMar 27, 201310 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 15, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 15, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 14, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 15, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing SUMMIT MEDICAL LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check SUMMIT MEDICAL LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Summit Medical, LLC
Location
Saint Paul, MN (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.