Lucero Medical LLC FDA Inspection Record
Lucero Medical LLC is an FDA-inspected facility (FEI 3008775608) located in Richfield, OH, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Sep 7, 2022: No Action Indicated (NAI)
- Inspections on record
- 4 (1 OAI, 1 VAI, 2 NAI)
- Last OAI inspection
- Nov 22, 2011
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Mar 21, 2012. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Jul 15, 2012
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 7, 2022 | No Action Indicated (NAI) | Design history file (21 CFR 820.30(j)) Lack of or inadequate procedures for labeling (21 CFR 820.120) Quality Audit/Reaudit - conducted (21 CFR 820.22) Supplier oversight (21 CFR 820.50(a)(2)) and 1 more |
| Sep 12, 2013 | No Action Indicated (NAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design validation - Risk analysis (21 CFR 820.30(g)) Design input - documentation (21 CFR 820.30(c)) Design validation - production units (21 CFR 820.30(g)) and 1 more |
| Jun 8, 2012 | Voluntary Action Indicated (VAI) | Design output - Lack of or inadequate procedures (21 CFR 820.30(d)) Design validation - Risk analysis (21 CFR 820.30(g)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Nonconforming product control (21 CFR 820.90(a)) and 3 more |
| Nov 22, 2011 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Quality System Procedures (21 CFR 820.20(e)) Design control - no procedures (21 CFR 820.30(a)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) and 6 more |
Warning letters
- Mar 21, 2012: Warning letter listed on FDA's Data Dashboard. View compliance record (269897).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 15, 2012 | Class II | Z-2368-2012 | Enduramesh Corpectomy Spacer System
Product Usage: The Enduramesh is a vertebral body… The firm voluntarily recalled the device after they discovered that the sterilization recommendations on the product insert did not pass… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2368-2012, this FEI as the recalling firm | Jul 15, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 8, 2012 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Mar 21, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Nov 22, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 27, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 27, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 27, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 27, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Lucero Medical LLC as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Lucero Medical LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Facility details
- Company
- Lucero Medical LLC
- Location
- Richfield, OH (US)
- Source
- inspections