DICOM Grid (Dba Ambra Health) FDA Inspection Record
DICOM Grid (Dba Ambra Health) is an FDA-registered facility (FEI 3008776294) located in Raleigh, NC, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jul 27, 2022: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (2 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Nov 18, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 27, 2022 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Review and evaluation for investigation (21 CFR 820.198(b)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) Justification for not reporting (21 CFR 806.20(b)(4)) |
| Sep 25, 2019 | Voluntary Action Indicated (VAI) | Design review - documentation (21 CFR 820.30(e)) Software validation for automated processes (21 CFR 820.70(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Design verification - documentation (21 CFR 820.30(f)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Nov 18, 2025 | Class II | Z-0941-2026 | Intelerad InteleShare software, with ProViewer component Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of… |
| Jun 23, 2022 | Class II | Z-1818-2022 | Ambra PACS, Ambra ProViewer. Software for use as a primary diagnostic and analysis tool… A race condition between the storage system and services database has the potential to revert edits made to patient information upon first… |
| Dec 16, 2016 | Class II | Z-1045-2017 | Ambra PACS
UDI: +AMBRHEALTHSOLUTIONS0/$$+7_V3.16.13.0R
Software Version Number… A software error caused the window/level to become the same in one series regardless if the image had different levels; image results have a… |
FDA record score
13 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0941-2026, this FEI as the recalling firm | Nov 18, 2025 | 4 x 1 | 4 |
| Inspection classified VAI | Jul 27, 2022 | 6 x 0.6 | 3.6 |
| Class II recall Z-1818-2022, this FEI as the recalling firm | Jun 23, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Sep 25, 2019 | 6 x 0.25 | 1.5 |
| Class II recall Z-1045-2017, this FEI as the recalling firm | Dec 16, 2016 | 4 x 0.25 | 1 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 16, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing DICOM Grid (Dba Ambra Health) as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check DICOM Grid (Dba Ambra Health) yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- DICOM Grid, Inc.
- Location
- Raleigh, NC (US)
- Source