Oxus America, Inc. FDA Inspection Record
Oxus America, Inc. is an FDA-registered facility (FEI 3008841036) located in Orion Township, MI, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Oct 10, 2019: No Action Indicated (NAI)
- Inspections on record
- 4 (2 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Jan 14, 2013
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 10, 2019 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Jan 19, 2016 | Voluntary Action Indicated (VAI) | Design control - no procedures (21 CFR 820.30(a)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) and 1 more |
| Sep 20, 2013 | No Action Indicated (NAI) | DMR - not or inadequately maintained (21 CFR 820.181) |
| Aug 1, 2011 | Voluntary Action Indicated (VAI) | Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Supplier notification of changes (21 CFR 820.50(b)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jan 14, 2013 | Class II | Z-0768-2013 | Li-Ion battery used in the Oxus Portable Oxygen Concentrator (POC), RS-00400.
Also… Due to past battery inventory storage conditions and deficiencies in the maintenance procedure, it was possible that some batteries were… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
3 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jan 19, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0768-2013, this FEI as the recalling firm | Jan 14, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 1, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Oxus America, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Oxus America, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Facility details
- Company
- Oxus, Inc.
- Location
- Orion Township, MI (US)
- Source
- device