Transfer Technology, Inc. FDA Inspection Record
Transfer Technology, Inc. is an FDA-registered facility (FEI 3009494532) located in Huntington Beach, CA, US. Registered operations: LABEL; MANUFACTURE; PACK; RELABEL; REPACK. RegDataLab tracks 5 FDA inspections at this facility.
Reviewing Transfer Technology, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Transfer Technology, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
What FDA found at this site
- Latest inspection
- Apr 30, 2025: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (4 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 30, 2025 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Sampling plans (21 CFR 820.250(b)) |
| Apr 28, 2023 | Voluntary Action Indicated (VAI) | Procedures not in writing, fully followed (21 CFR 211.22(d)) |
| Apr 13, 2021 | Voluntary Action Indicated (VAI) | DHR - not or inadequately maintained (21 CFR 820.184) Design control - no procedures (21 CFR 820.30(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of or inadequate procedures for labeling (21 CFR 820.120) and 3 more |
| Oct 11, 2019 | No Action Indicated (NAI) | DHR - not or inadequately maintained (21 CFR 820.184) Procedures not adequately established or maintained (21 CFR 820.40) Design input - documentation (21 CFR 820.30(c)) Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h)) and 8 more |
| Sep 17, 2013 | Voluntary Action Indicated (VAI) | Design plans - Lack of or inadequate (21 CFR 820.30(b)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h)) DMR production process specifications (21 CFR 820.181(b)) and 8 more |
Facility details
- Company
- Transfer Technology, Inc.
- Location
- Huntington Beach, CA (US)
- Registration Expires
- Dec 31, 2026
- Operations
- LABEL; MANUFACTURE; PACK; RELABEL; REPACK
- Source
- decrs