Medartis Inc. FDA Inspection Record

FEI 3009540749 · Inspections, Form 483s, warning letters and import refusals

Medartis Inc. is an FDA-registered facility (FEI 3009540749) located in Warsaw, IN, US. RegDataLab tracks 2 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Jun 3, 2026: Voluntary Action Indicated (VAI)
Inspections on record
2 (2 VAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
8, latest Apr 12, 2024

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jun 3, 2026Voluntary Action Indicated (VAI)Methods for monitoring, measurement, analysis, and improvement not determined (ISO 13485:2016 Clause 8.1)
Information gathered in feedback process not used as input into risk management (ISO 13485:2016 Clause 8.2.1)
Procedures for analysis of data not documented (ISO 13485:2016 Clause 8.4)
Sep 3, 2021Voluntary Action Indicated (VAI)Nonconforming product control (21 CFR 820.90(a))
Investigation of device failures (21 CFR 820.198(c))
Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Apr 12, 2024Class IIZ-1821-2024APTUS 2.8 TriLock Screw 22mm, HD7, 1/Pkg. Metal bone fixation fastener used with the APTUS…
Screw is 16mm long instead of 22mm long.
Feb 10, 2022Class IIZ-0910-2022Nextra Hammertoe Correction System 4.5 Middle 3.2 Proximal Kit SKU: NX-4532K… and 4 more products
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra…
Jan 13, 2022Class IIZ-0739-2022Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone reconstruction… and 1 more product
During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate…
Mar 26, 2021Class IIZ-1500-2021InCore Lapidus Disposable Kit 28mm Left, Part Number: IC-LAP-28L - Product Usage: Intended… and 1 more product
Product could potentially contain an undersized inner diameter of the 40mm drill bushing which result in not allowing the drill bit to pass…
Feb 12, 2020Class IIZ-1458-2020Nextra Hammertoe Correction System-4.5MM Middle, 3.2MM Proximal, SS, Sterile. REF…
Proximal end of the driver in the affected product may be oversized in varying degrees, potentially making it difficult to connect with and…
Dec 16, 2019Class IIZ-1127-2020Stratum Lapidus Plate - Small Rt, Part No. STRM-LAP-SMR, UDI 00817701025467. The product…
The product contains a different plate than indicated by the package labeling. The product is labeled as a right plate but contains a left…
Nov 11, 2019Class IIZ-0776-2020Stratum MDS Screw 3.5x46mm ST Part Number: STRM-MDS-3546ST and 2 more products
Components 2mm shorter than indicated on the package label
Oct 9, 2019Class IIZ-0534-2020Stratum Foot Plating System-NL LP Screw 3.5x50mm ST, sterile. Catalog Number…
Product labeled as STRM-NL-3550ST, NL LP Screw 3.5x50mm ST, lot M1151 may contain the incorrect screw type and length

FDA record score

20 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 17, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIJun 3, 20266 x 16
Class II recall Z-1821-2024, this FEI as the recalling firmApr 12, 20244 x 0.62.4
Class II recall Z-0910-2022, this FEI as the recalling firmFeb 10, 20224 x 0.62.4
Class II recall Z-0739-2022, this FEI as the recalling firmJan 13, 20224 x 0.62.4
Inspection classified VAISep 3, 20216 x 0.251.5
Class II recall Z-1500-2021, this FEI as the recalling firmMar 26, 20214 x 0.251
Class II recall Z-1458-2020, this FEI as the recalling firmFeb 12, 20204 x 0.251
Class II recall Z-1127-2020, this FEI as the recalling firmDec 16, 20194 x 0.251
Class II recall Z-0776-2020, this FEI as the recalling firmNov 11, 20194 x 0.251
Class II recall Z-0534-2020, this FEI as the recalling firmOct 9, 20194 x 0.251

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 17, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 17, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 17, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 17, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 17, 2026Newest record Nov 16, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing Medartis Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Medartis Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Medartis Inc.
Location
Warsaw, IN (US)
Source
device