Conformis, Inc. FDA Inspection Record

FEI 3009844603 · Inspections, Form 483s, warning letters and import refusals

Conformis, Inc. is an FDA-inspected facility (FEI 3009844603) located in Billerica, MA, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Mar 9, 2023: Voluntary Action Indicated (VAI)
Inspections on record
5 (3 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
19, latest Mar 11, 2024

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 9, 2023Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design input - Lack of or inadequate procedures (21 CFR 820.30(c))
Lack of or inadequate procedures (21 CFR 820.100(a))
Investigation of device failures (21 CFR 820.198(c))
Sep 20, 2019Voluntary Action Indicated (VAI)Design validation - Risk analysis (21 CFR 820.30(g))
Design input - documentation (21 CFR 820.30(c))
Sep 7, 2018Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Documentation (21 CFR 820.75(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 2 more
Jun 12, 2015No Action Indicated (NAI)No citations listed
Mar 18, 2013No Action Indicated (NAI)No citations listed

Recalls

Showing the 10 most recent of 19 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Mar 11, 2024Class IIZ-1591-2024Identity Imprint PS Tibial Tray Size 4: Lot 540287
Size 5 Tibial Tray from Lot 508287 was packaged in a Size 4 box from Lot 540287. It was confirmed that Lot 540287 and Lot 508287 were…
Jun 16, 2023Class IIZ-2405-2023CONFORMIS HIP-SCREW-6.5MM X 25MM, STERILE EO
30mm screws were labeled as 25mm screws.
Nov 14, 2022Class IIZ-1133-2023iTotal Identity Impactor Handle, Model No. 1080-212/ED-07827 REV AE
Loctite thread locker (epoxy) may not have been applied to the screw that is part of the handle. In the absence of epoxy, the screw may…
Feb 22, 2022Class IIZ-0899-2022iTotal Identity Posterior Stabilised (PS) Knee Replacement System Model Number…
Incorrect tibial base plate implant packaged in kit
Sep 21, 2020Class IIZ-0475-2021iTotal Hip Replacement System, Model number HAA-050-D020-020102 The Conformis Hip…
Incorrect hip components were provided in kits.
Jul 20, 2020Class IIZ-2849-2020Conformis iTotal Cruciate Retaining (CR) Knee Replacement System- Orthopedic Implant Cat…
Outer box labels for kit 0469247 correct, however the boxes contained the implants and jigs for 0468920
Feb 3, 2020Class IIZ-1627-2020iTotal CR Knee Replacement System, Model No. TCR-121-1111-010102, (ITOTAL CR IPOLY XE… and 1 more product
A limited number of tibial trays were labeled with an incorrect serial number and were delivered to the wrong customer accounts.
Oct 25, 2019Class IIZ-0672-2020iTotal CR Knee Replacement System Model Number: TCR-111-1111-010102 - Product Usage: The…
Medial and lateral polyethylene inserts did not securely fit into the cobalt chrome tray implant.
Aug 22, 2019Class IIZ-0063-2020iTotal Hip Replacement System, Model HBS-033-0015-020101, Size 15 Hip-Patient Specific… and 1 more product
Two femoral stems, sizes 13 and 15, were erroneously switched and packed in each other's packaging.
Mar 29, 2019Class IIZ-1775-2019Conformis iTotal Hip Replacement System: HBS-033-00xx-010101, (HIP-PATIENT SPECIFIC STEM… and 1 more product
Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head…

FDA record score

26 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

22 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIMar 9, 20236 x 0.63.6
Class II recall Z-1591-2024, this FEI as the recalling firmMar 11, 20244 x 0.62.4
Class II recall Z-2405-2023, this FEI as the recalling firmJun 16, 20234 x 0.62.4
Class II recall Z-1133-2023, this FEI as the recalling firmNov 14, 20224 x 0.62.4
Class II recall Z-0899-2022, this FEI as the recalling firmFeb 22, 20224 x 0.62.4
Inspection classified VAISep 20, 20196 x 0.251.5
Inspection classified VAISep 7, 20186 x 0.251.5
Class II recall Z-0475-2021, this FEI as the recalling firmSep 21, 20204 x 0.251
Class II recall Z-2849-2020, this FEI as the recalling firmJul 20, 20204 x 0.251
Class II recall Z-1627-2020, this FEI as the recalling firmFeb 3, 20204 x 0.251
Class II recall Z-0672-2020, this FEI as the recalling firmOct 25, 20194 x 0.251
Class II recall Z-0063-2020, this FEI as the recalling firmAug 22, 20194 x 0.251
Class II recall Z-1775-2019, this FEI as the recalling firmMar 29, 20194 x 0.251
Class II recall Z-1079-2019, this FEI as the recalling firmJan 4, 20194 x 0.251
Class II recall Z-1108-2019, this FEI as the recalling firmDec 3, 20184 x 0.251
Class II recall Z-0731-2019, this FEI as the recalling firmOct 16, 20184 x 0.251
Class II recall Z-0600-2018, this FEI as the recalling firmJul 24, 20174 x 0.251
Class II recall Z-0991-2017, this FEI as the recalling firmOct 31, 201610 years or older, so it adds no points0
Class II recall Z-0658-2017, this FEI as the recalling firmOct 18, 201610 years or older, so it adds no points0
Class II recall Z-1070-2016, this FEI as the recalling firmJan 28, 201610 years or older, so it adds no points0
Class II recall Z-0138-2016, this FEI as the recalling firmAug 31, 201510 years or older, so it adds no points0
Class II recall Z-0581-2015, this FEI as the recalling firmNov 19, 201410 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Conformis, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Conformis, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Conformis, Inc.
Location
Billerica, MA (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.