Spacelabs Healthcare Inc. FDA Inspection Record
Spacelabs Healthcare Inc. is an FDA-registered facility (FEI 3010157426) located in Snoqualmie, WA, US. RegDataLab tracks 7 FDA inspections and 2 compliance actions at this facility.
What FDA found at this site
- Latest inspection
- Sep 15, 2022: Voluntary Action Indicated (VAI)
- Inspections on record
- 7 (1 OAI, 5 VAI, 1 NAI)
- Last OAI inspection
- Apr 2, 2014
- Warning letters
- 2 listed by FDA for this FEI, latest Aug 5, 2014
- Recalls as the recalling firm
- 28, latest Jun 4, 2026
- Import refusals
- 2
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 15, 2022 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a)) |
| Jun 1, 2018 | No Action Indicated (NAI) | No citations listed |
| Mar 9, 2017 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) |
| Jul 31, 2015 | Voluntary Action Indicated (VAI) | Design validation - production units (21 CFR 820.30(g)) Not approved or obsolete document retrieval (21 CFR 820.40(a)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Reporting Information Reasonably Known (21 CFR 803.50(b)(1)) |
| Apr 2, 2014 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design verification - output does not meet input requirement (21 CFR 820.30(f)) Nonconforming product control (21 CFR 820.90(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 4 more |
| Apr 18, 2012 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Documentation (21 CFR 820.80(e)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Mar 12, 2010 | Voluntary Action Indicated (VAI) | Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Design verification acceptance criteria (21 CFR 820.30(f)) Lack of or inadequate procedures (21 CFR 820.100(a)) Not approved or obsolete document retrieval (21 CFR 820.40(a)) and 7 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 4, 2026 | Class II | Z-2916-2026 | Brand Name: Xhibit
Product Name: Xhibit Central Station
Model/Catalog Number… Due to off-the-shelf display port component that has audio/video failures |
| Mar 26, 2026 | Class II | Z-2052-2026 | Brand Name: Spacelabs Ultraview SL Command Module
Product Name: SL Command… Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester). |
| Feb 27, 2026 | Class II | Z-1793-2026 | Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early… |
| Sep 5, 2025 | Class II | Z-0167-2026 | Brand Name: Branding is not applied to this product.
Product Name: DVI display… and 2 more products Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may… |
| Nov 25, 2024 | Class II | Z-0809-2025 | Xhibit Telemetry Receiver, Model 96280, with Service Manual.
Intended to provide… Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting… |
| Nov 25, 2024 | Class II | Z-0810-2025 | Xhibit Telemetry Receiver with Software, Model 96280.
Intended to provide monitoring… Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting… |
| Apr 12, 2023 | Class II | Z-1503-2023 | Ultraview SL (UVSL) Command Module, Model 91496 Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will… |
| May 31, 2022 | Class II | Z-1513-2022 | Xhibit Telemetry Receiver, Model: 96280 Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following… |
| Aug 30, 2019 | Class II | Z-2261-2020 | Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians… The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient… |
| Jan 28, 2019 | Class II | Z-2035-2019 | Spacelabs Healthcare Smart Disclosure System, Model 92810, a component of the Intesys… Several reports were received that patient records were printed with the correct patient demographics but containing another patient's… |
FDA record score
38 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
36 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2916-2026, this FEI as the recalling firm | Jun 4, 2026 | 4 x 1 | 4 |
| Class II recall Z-2052-2026, this FEI as the recalling firm | Mar 26, 2026 | 4 x 1 | 4 |
| Class II recall Z-1793-2026, this FEI as the recalling firm | Feb 27, 2026 | 4 x 1 | 4 |
| Class II recall Z-0167-2026, this FEI as the recalling firm | Sep 5, 2025 | 4 x 1 | 4 |
| Class II recall Z-0809-2025, this FEI as the recalling firm | Nov 25, 2024 | 4 x 1 | 4 |
| Class II recall Z-0810-2025, this FEI as the recalling firm | Nov 25, 2024 | 4 x 1 | 4 |
| Inspection classified VAI | Sep 15, 2022 | 6 x 0.6 | 3.6 |
| Class II recall Z-1503-2023, this FEI as the recalling firm | Apr 12, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1513-2022, this FEI as the recalling firm | May 31, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Mar 9, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-2261-2020, this FEI as the recalling firm | Aug 30, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2035-2019, this FEI as the recalling firm | Jan 28, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0532-2018, this FEI as the recalling firm | Jan 17, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2093-2017, this FEI as the recalling firm | Apr 14, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0332-2017, this FEI as the recalling firm | Oct 18, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2885-2016, this FEI as the recalling firm | Aug 25, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2343-2016, this FEI as the recalling firm | Jul 1, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1502-2016, this FEI as the recalling firm | Mar 28, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1092-2016, this FEI as the recalling firm | Feb 26, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0172-2016, this FEI as the recalling firm | Sep 3, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 31, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1608-2015, this FEI as the recalling firm | May 8, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1577-2015, this FEI as the recalling firm | Apr 21, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1542-2015, this FEI as the recalling firm | Apr 17, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1144-2015, this FEI as the recalling firm | Jan 28, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2707-2014, this FEI as the recalling firm | Aug 26, 2014 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Aug 5, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2586-2014, this FEI as the recalling firm | Jul 2, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2077-2014, this FEI as the recalling firm | Jun 18, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1852-2014, this FEI as the recalling firm | Jun 6, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1824-2014, this FEI as the recalling firm | May 9, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1705-2014, this FEI as the recalling firm | May 8, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Apr 2, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 18, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 12, 2010 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Nov 3, 2008 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |