IMRIS Imaging, Inc. FDA Inspection Record

FEI 3010326005 · Inspections, Form 483s, warning letters and import refusals

IMRIS Imaging, Inc. is an FDA-registered facility (FEI 3010326005) located in Chaska, MN, US. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Oct 1, 2019: Voluntary Action Indicated (VAI)
Inspections on record
3 (2 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
16, latest Oct 16, 2026

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Oct 1, 2019Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Documentation (21 CFR 820.100(b))
Report of assembly submitted within 15 days (21 CFR 1020.30(d)(1))
Jan 26, 2018Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))
Report of risk to health (21 CFR 806.10(a)(1))
and 1 more
Feb 4, 2015No Action Indicated (NAI)No citations listed

Recalls

Showing the 10 most recent of 16 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Oct 16, 2026Class IIZ-0441-2027IMRIS Operating Suite iMRX System, Model Number 105828-000, IMRIS Product Description… and 4 more products
Operating suite contains TruSystem Surgical Tables which are being recalled by Baxter.
Aug 6, 2026Class IIZ-3033-2026ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned… and 1 more product
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning…
Dec 8, 2025Class IIZ-1034-2026Brand Name: HFD100 Product Name: Head Fixation Device HFD100 Model/Catalog Number…
Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision…
Sep 30, 2025Class IZ-0161-2026IMRIS Operating Suite, Neuro III-SV iMRI System, iMRX System, containing Siemens 3… and 3 more products
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating…
Feb 17, 2022Class IIZ-0934-2022Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of…
Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may…
Aug 25, 2021Class IIZ-2584-2021Operating room table - Product Usage: intended for use during diagnostic examinations… and 2 more products
ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware…
Nov 26, 2019Class IIZ-0762-2020iMRX MRI and Angiography System, part number 10094137 and 10094141, a…
There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator…
Jun 26, 2019Class IIZ-0163-2020IMRIS MR/X-ray head fixation device, Model HFD200
Breaking of the HFD200 Drive Knob force indicator due to over-torqueing of the drive knob beyond the maximum allowable pinning force…
Nov 6, 2018Class IIZ-0645-2019IMRIS T2X table, Part Numbers: (a)108000-000, (b)108000-600, (c)109682-000, (d)… and 3 more products
Table may drift in the roll position while in use
Apr 4, 2018Class IIZ-1645-2018IMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS intraoperative MRI systems…
The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our…

FDA record score

41 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 17, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

18 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-0161-2026, this FEI as the recalling firmSep 30, 202515 x 115
Class II recall Z-0441-2027, this FEI as the recalling firmOct 16, 20264 x 14
Class II recall Z-3033-2026, this FEI as the recalling firmAug 6, 20264 x 14
Class II recall Z-1034-2026, this FEI as the recalling firmDec 8, 20254 x 14
Class II recall Z-0934-2022, this FEI as the recalling firmFeb 17, 20224 x 0.62.4
Inspection classified VAIOct 1, 20196 x 0.251.5
Inspection classified VAIJan 26, 20186 x 0.251.5
Class II recall Z-2584-2021, this FEI as the recalling firmAug 25, 20214 x 0.251
Class II recall Z-0762-2020, this FEI as the recalling firmNov 26, 20194 x 0.251
Class II recall Z-0163-2020, this FEI as the recalling firmJun 26, 20194 x 0.251
Class II recall Z-0645-2019, this FEI as the recalling firmNov 6, 20184 x 0.251
Class II recall Z-1645-2018, this FEI as the recalling firmApr 4, 20184 x 0.251
Class II recall Z-1217-2018, this FEI as the recalling firmFeb 23, 20184 x 0.251
Class II recall Z-1263-2018, this FEI as the recalling firmFeb 23, 20184 x 0.251
Class II recall Z-0975-2018, this FEI as the recalling firmFeb 20, 20184 x 0.251
Class II recall Z-3211-2017, this FEI as the recalling firmAug 14, 20174 x 0.251
Class II recall Z-1098-2016, this FEI as the recalling firmFeb 10, 201610 years or older, so it adds no points0
Class II recall Z-0144-2015, this FEI as the recalling firmOct 1, 201410 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 16, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 16, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 16, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 16, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing IMRIS Imaging, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check IMRIS Imaging, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
IMRIS Imaging Inc
Location
Chaska, MN (US)
Source
device