IMRIS Imaging, Inc. FDA Inspection Record
IMRIS Imaging, Inc. is an FDA-registered facility (FEI 3010326005) located in Chaska, MN, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Oct 1, 2019: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 16, latest Oct 16, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 1, 2019 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Documentation (21 CFR 820.100(b)) Report of assembly submitted within 15 days (21 CFR 1020.30(d)(1)) |
| Jan 26, 2018 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) Report of risk to health (21 CFR 806.10(a)(1)) and 1 more |
| Feb 4, 2015 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 16, 2026 | Class II | Z-0441-2027 | IMRIS Operating Suite iMRX System, Model Number 105828-000, IMRIS Product Description… and 4 more products Operating suite contains TruSystem Surgical Tables which are being recalled by Baxter. |
| Aug 6, 2026 | Class II | Z-3033-2026 | ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned… and 1 more product The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning… |
| Dec 8, 2025 | Class II | Z-1034-2026 | Brand Name: HFD100
Product Name: Head Fixation Device HFD100
Model/Catalog Number… Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision… |
| Sep 30, 2025 | Class I | Z-0161-2026 | IMRIS Operating Suite, Neuro III-SV iMRI System, iMRX System, containing Siemens 3… and 3 more products There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating… |
| Feb 17, 2022 | Class II | Z-0934-2022 | Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of… Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may… |
| Aug 25, 2021 | Class II | Z-2584-2021 | Operating room table - Product Usage: intended for use during diagnostic examinations… and 2 more products ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware… |
| Nov 26, 2019 | Class II | Z-0762-2020 | iMRX MRI and Angiography System, part number 10094137 and 10094141, a… There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator… |
| Jun 26, 2019 | Class II | Z-0163-2020 | IMRIS MR/X-ray head fixation device, Model HFD200 Breaking of the HFD200 Drive Knob force indicator due to over-torqueing of the drive knob beyond the maximum allowable pinning force… |
| Nov 6, 2018 | Class II | Z-0645-2019 | IMRIS T2X table, Part Numbers: (a)108000-000, (b)108000-600, (c)109682-000, (d)… and 3 more products Table may drift in the roll position while in use |
| Apr 4, 2018 | Class II | Z-1645-2018 | IMRIS IMRI 1.5T AND 3T S System
Product Usage:
The IMRIS intraoperative MRI systems… The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our… |
FDA record score
41 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
18 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-0161-2026, this FEI as the recalling firm | Sep 30, 2025 | 15 x 1 | 15 |
| Class II recall Z-0441-2027, this FEI as the recalling firm | Oct 16, 2026 | 4 x 1 | 4 |
| Class II recall Z-3033-2026, this FEI as the recalling firm | Aug 6, 2026 | 4 x 1 | 4 |
| Class II recall Z-1034-2026, this FEI as the recalling firm | Dec 8, 2025 | 4 x 1 | 4 |
| Class II recall Z-0934-2022, this FEI as the recalling firm | Feb 17, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Oct 1, 2019 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Jan 26, 2018 | 6 x 0.25 | 1.5 |
| Class II recall Z-2584-2021, this FEI as the recalling firm | Aug 25, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0762-2020, this FEI as the recalling firm | Nov 26, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0163-2020, this FEI as the recalling firm | Jun 26, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0645-2019, this FEI as the recalling firm | Nov 6, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1645-2018, this FEI as the recalling firm | Apr 4, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1217-2018, this FEI as the recalling firm | Feb 23, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1263-2018, this FEI as the recalling firm | Feb 23, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0975-2018, this FEI as the recalling firm | Feb 20, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-3211-2017, this FEI as the recalling firm | Aug 14, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1098-2016, this FEI as the recalling firm | Feb 10, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0144-2015, this FEI as the recalling firm | Oct 1, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 16, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing IMRIS Imaging, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check IMRIS Imaging, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- IMRIS Imaging Inc
- Location
- Chaska, MN (US)
- Source